overwrite any previously stored programs in the selected dip switch pattern. 

Choose a new program number and repeat these steps.

Marely switching program numbers will not update the program to what is
currently stored  on  the  new  number.  It  will  maintain  the  most  recently
stored  program until changes are made and STORE button is pressed.  

Selecting 

READ (for current settings) 

will display what program is 

currently stored for a particular dip switch patten in WATSON.

All programs that are stored in WATSON can be accessed by selecting the
program number of a dip switch pattern and pressing the 

READ 

button

.

If no program is stored for the dip switch combination the program will 
say that the current program is empty

Close the application and disconnect WATSON from the computer once 

the programming is completed.

7 | 

P a g e

WATSON

 Manual Ver 1.0

Содержание PCR

Страница 1: ......

Страница 2: ...te capacity cooling fan producing 2 0 5 degrees sec cooling ramp time The heated lid prevents water condensation during sample heating step Power is delivered to WATSON by a USB type C port Either USB...

Страница 3: ...improperly handled Safety Compliance This instrument complies with applicable clauses of the requirement for the following safety standards Applicable clauses of IEC 61010 1 2001 Safety Requirements...

Страница 4: ...e the machine into the STANDBY mode for accepting samples The blue LED is for power and the green LED is for the STANDBY mode Setup the reagent system for amplifying a particular nucleic acid target b...

Страница 5: ...h the run It is also a burn hazard After the program is completed the LEDs will switch back to solid blue and green Solid blue and green indicates that the program is completed and is in the STANDBY m...

Страница 6: ...RNA into DNA For converting RNA to DNA this stage will require two steps one for the enzyme reaction and the other for destroying the enzyme by heat Otherwise only one step is required during the pre...

Страница 7: ...ton is pressed Selecting READ for current settings will display what program is currently stored for a particular dip switch patten in WATSON All programs that are stored in WATSON can be accessed by...

Страница 8: ...the WATSON with assay tubes and press RUN The top right corner of the interface shows the lid and plate temperature The top left corner shows the estimated run time for the PCR program The red and bl...

Страница 9: ...ng Red Device is cooling down after PCR has completed Blinking Green and Red Device has encountered a temperature problem please contact technical support Dip Switch Settings Figure 4 Dip Switch with...

Страница 10: ...ll 1 Up 1 Down 2 3 4 2 Up 2 Down 1 3 4 3 Up 1 2 Down 3 4 4 Up 3 Down 1 2 4 5 Up 1 3 Down 2 4 6 Up 2 3 Down 1 4 7 Up 1 2 3 Down 4 8 Up 4 Down 1 2 3 9 Up 1 4 Down 2 3 10 Up 2 4 Down 1 3 11 Up 1 2 4 Down...

Страница 11: ...N from power and connection to the computer Quit the software and reconnect WATSON to power and to the computer and reopen the software Maintenance Prior to doing any maintenance please unplug power s...

Страница 12: ...ple Format 4 wells plate PCR tubes 200 ul capacity Thermocycling System Resistive Heating Forced Air Cooling Programmable Plate Temperature range 30 105 C Programmable Lid Temperature range 30 105 C T...

Страница 13: ...TNESS FOR A PARTICULAR PURPOSE EITHER EXPRESSED OR IMPLIED RELATED TO THE PERFORMANCE OF THIS PRODUCT BUYER S SOLE REMEDY FOR DEFECTIVE MATERIALS IS REPLACEMENT IN NO EVENT SHALL MICROBIO BE LIABLE FO...

Отзывы: