Download the WATSON software into your computer from the WATSON 

website (www.WATSONPCR.com).

Connect WATSON to power supply via USB-C.

Connect WATSON to a computer via a Micro USB-B cable using the micro

USB port. The LEDs should be solid blue and green.

Open the WATSON software and 

verify on the bottom left of the Main 

interface

 window that the device is connected.

Choose which program to alter by changing the 

Program Number 

or by 

altering the dip switch patterns on WATSON or on the dip switch icon on the main
program interface.

Alter temperature and duration of each step by selecting the desired box 

and changing the numbers. Programmable temperature for the plate range from 
30°C to 100°C.

The planned program will be shown on the graph on the main interface 

and the plot will update when alterations are made to time and temperature 
parameters. The.

Prep Stage: Step 1 and 2:  Mostly needed for denaturing DNA 

templates or converting RNA into DNA

. For converting RNA to DNA, this 

stage will require two steps, one for the enzyme reaction and the other for 
destroying the enzyme by heat. Otherwise, only one step is required during the 
prep stage. 

Run Stage: Step 3, 4,  and 5:  This stage 

is the main cycling program to 

amplify samples. The run stage can be either two steps or three steps PCR.  “

x” 

symbol 

delete steps and “

+” symbol 

adds steps. Most PCR can be completed 

by a two stage reaction. 

Hold Stage: Step 6 and 7: 

This stage is to ensure that the device cools 

off to room temperature or remains at a set desired temperature.

Number of Cycles 

sets the amount of cycles WATSON will undergo of 

during the Run Stage. Typical cycle numbers are 35-40. 

Lid Temperature 

sets the temperature that will be maintained by heated 

lid plate throughout the duration of the program. Temperature range is between 
30°C and 105°C. A heated lid will prevent water condensation into the lid of the 
test tube which is undesirable. 

After all steps have their temperature and time parameters set, select 

STORE button 

to upload the program to WATSON. This process will also will 

6 | 

P a g e

WATSON

 Manual Ver 1.0

Содержание PCR

Страница 1: ......

Страница 2: ...te capacity cooling fan producing 2 0 5 degrees sec cooling ramp time The heated lid prevents water condensation during sample heating step Power is delivered to WATSON by a USB type C port Either USB...

Страница 3: ...improperly handled Safety Compliance This instrument complies with applicable clauses of the requirement for the following safety standards Applicable clauses of IEC 61010 1 2001 Safety Requirements...

Страница 4: ...e the machine into the STANDBY mode for accepting samples The blue LED is for power and the green LED is for the STANDBY mode Setup the reagent system for amplifying a particular nucleic acid target b...

Страница 5: ...h the run It is also a burn hazard After the program is completed the LEDs will switch back to solid blue and green Solid blue and green indicates that the program is completed and is in the STANDBY m...

Страница 6: ...RNA into DNA For converting RNA to DNA this stage will require two steps one for the enzyme reaction and the other for destroying the enzyme by heat Otherwise only one step is required during the pre...

Страница 7: ...ton is pressed Selecting READ for current settings will display what program is currently stored for a particular dip switch patten in WATSON All programs that are stored in WATSON can be accessed by...

Страница 8: ...the WATSON with assay tubes and press RUN The top right corner of the interface shows the lid and plate temperature The top left corner shows the estimated run time for the PCR program The red and bl...

Страница 9: ...ng Red Device is cooling down after PCR has completed Blinking Green and Red Device has encountered a temperature problem please contact technical support Dip Switch Settings Figure 4 Dip Switch with...

Страница 10: ...ll 1 Up 1 Down 2 3 4 2 Up 2 Down 1 3 4 3 Up 1 2 Down 3 4 4 Up 3 Down 1 2 4 5 Up 1 3 Down 2 4 6 Up 2 3 Down 1 4 7 Up 1 2 3 Down 4 8 Up 4 Down 1 2 3 9 Up 1 4 Down 2 3 10 Up 2 4 Down 1 3 11 Up 1 2 4 Down...

Страница 11: ...N from power and connection to the computer Quit the software and reconnect WATSON to power and to the computer and reopen the software Maintenance Prior to doing any maintenance please unplug power s...

Страница 12: ...ple Format 4 wells plate PCR tubes 200 ul capacity Thermocycling System Resistive Heating Forced Air Cooling Programmable Plate Temperature range 30 105 C Programmable Lid Temperature range 30 105 C T...

Страница 13: ...TNESS FOR A PARTICULAR PURPOSE EITHER EXPRESSED OR IMPLIED RELATED TO THE PERFORMANCE OF THIS PRODUCT BUYER S SOLE REMEDY FOR DEFECTIVE MATERIALS IS REPLACEMENT IN NO EVENT SHALL MICROBIO BE LIABLE FO...

Отзывы: