AC Power consumption
190VA
DC Supply voltage/frequency
(12/24VDC power option)
12-24V DC
DC Power consumption (12/24VDC
power option)
150W
Installation category (overvoltage
category)
II
±10% of nominal voltage. Maximum
voltage fluctuation
An electrical mains supply is required
along with cable connections to the best
practice of noise immunity
Enclosure rating
IP66 to BS EN 60529
NEMA 4X to NEMA 250*
Operating temperature range
4°C to 45°C, 41°F to 113°F
Storage temperature range
-20°C to 70°C, -4°F to 158°F
Maximum altitude
2,000m, 6,560ft
Humidity (non-condensing)
80% up to 31°C, 88°F, decreasing linearly
to 50% at 40°C, 104°F
Pollution degree
2
Noise
<70dB(A) at 1m
*Requires the fitting of the HMI protective cover.
6.2
Standards (AC mains power supply)
EC Harmonised
standards
Safety requirements for electrical equipment for measurement,
control and laboratory use: BS EN 61010- 1 incorporating A2
Category 2, Pollution degree 2
Degrees of protection provided by enclosures (IP code): BS EN
60529 amendments 1 and 2
EN61326-1:2006 Electrical Equipment for measurement control
and laboratory use EMC requirements Part 1
m-qdos-en-02
16