Watson-Marlow Alitea 114 pumpheads: operating instructions

10

To change from the large tube to the small tube setting

Switch off the pump before changing the tube holder position. Use a pointed
device such as a ball-point pen to reposition the lower tube holders 

on both sides

of the pumphead.

Press down and slightly away from the front of the pumphead, as shown in the
first picture above.

Maintain the angled downward pressure and push away from the front of the
pumphead. The jaw clicks into a new position.

Release the pressure. The jaw rises into its correct alignment. If it does not
rise, repeat the procedure, being sure to maintain downward pressure until
release.

Adjust the tube holder on the other side of the pumphead in the same way.

To change from the small tube to the large tube setting

Carry out the procedure described above, but pushing towards the front of the
pumphead.

Lift the flip top until fully open.

Place the pointed device pointing
down into the small depression pic-
tured here.

Note

: The pictures on the previous page show the tube holders’ correct positions for

small and large tubing. If a tube holder is not vertical relative to the body of the
pumphead, it is wrongly positioned. Follow the instructions above to reposition it.

7.1 Tube holder repositioning

114DV, 114DVP

Содержание Alitea 114 Series

Страница 1: ...fications 4 3 1 Dimensions 4 4 Good pump installation practice 5 4 1 General recommendations 5 4 2 Do s and do not s 6 4 3 Troubleshooting 6 5 Pumphead mounting 7 6 Torque requirement 8 7 Tube holder...

Страница 2: ...Fabriksvag 29 31 120 33 Stockholm Sweden Telephone 46 8 556 556 00 Fax 46 8 556 556 25 The information in this user guide is believed to be correct at the time of publica tion However Watson Marlow Al...

Страница 3: ...Work Act 1974 There are moving parts inside the pumphead Before opening the flip top ensure that the following safety directions are followed Ensure that the pump is isolated from the mains power Ensu...

Страница 4: ...tinuous flow and up to 510 ml min intermittent flow Two tube holder positions to accept tube in bore sizes from 0 5mm to 4 8mm Models with occlusion settings for standard or high pressure operation Un...

Страница 5: ...r if the rotor rotates clockwise and that performance against pressure will be maximised if the rotor rotates counter clockwise Peristaltic pumps are self priming and self sealing against backflow No...

Страница 6: ...lse losses and pulsation in the pipeline This is especially important with viscous fluids and when connecting to rigid pipework Do site the pump at or just below the level of the fluid to be pumped if...

Страница 7: ...Mount the pumphead on its supporting plate through the holes indicated below using two Allen key headed M4 screws length equal to 5mm the thickness of the mount Tighten them to a maximum of 0 5Nm Driv...

Страница 8: ...he above figures x 2 Assess your application s torque requirement from the top of the arrows if all the parameters listed are those requiring more torque as shown in the graph below or to be sure to h...

Страница 9: ...and wandering across the rollers causing premature tube rupture With the larger bore tubes of 4 0mm and 4 8mm the outer position must be used to prevent the flow rate being excessively reduced For tub...

Страница 10: ...to its correct alignment If it does not rise repeat the procedure being sure to maintain downward pressure until release Adjust the tube holder on the other side of the pumphead in the same way To cha...

Страница 11: ...ck pressed against the pumphead inner wall The tube must not be twisted or stretched against the rollers Lower the flip top until it clicks into its fully closed position The track closes automaticall...

Страница 12: ...14 29 47 5 67 85 190 rpm 4 3 8 2 26 5 55 90 5 128 160 200 rpm 4 6 8 6 28 58 95 135 170 350 rpm 8 0 15 49 100 165 235 300 400 rpm 9 1 17 56 115 190 270 340 600 rpm 13 5 26 84 175 285 405 510 Flow rate...

Страница 13: ...e rotation optimizes the pump performance regarding tube life and required torque Clockwise rotation is recommended for low or no pressure applica tions Suction performance and flow are the same as fo...

Страница 14: ...2 0024 016 933 0024 016 913 A024 016 3 2 16 902 0032 016 933 0032 016 913 A032 016 4 0 933 0040 016 4 8 25 902 0048 016 933 0048 016 913 A048 016 mm Tygon E3603 Neoprene 0 8 13 920 0008 016 1 6 14 E36...

Страница 15: ...aint Gobain Corporation STA PURE PFL and STA PURE PCS are trademarks of W L Gore and Associates 12 Warning not to use pumps in patient connected applications Warning These products are not designed fo...

Отзывы: