Watson-Marlow Alitea 114 Series Скачать руководство пользователя страница 6

Watson-Marlow Alitea 114 pumpheads: operating instructions

6

4.2 Do’s and do not’s

Do not

build a pump into a tight location without adequate airflow around the pump.

Do

keep delivery and suction tubes as short and direct as possible - though ideally

not shorter than 0.5m - and follow the straightest route. Use bends of large radius:
at least four times the tubing diameter. Ensure that connecting pipework and fittings
are suitably rated to handle the predicted pipeline pressure. Try to avoid pipe reduc-
ers and lengths of smaller bore tubing than the pumphead section, 

particularly in

pipelines on the suction side

. Any valves in the pipeline (not usually needed with

a self-priming peristaltic pump) must not restrict the flow. Any valves in the flow line
must be open when the pump is running.

Do

use suction and delivery pipes equal to or larger than the bore of the tube in the

pumphead. When pumping viscous fluids use pipe runs with a bore several times
larger than the pump tube.

Do

ensure that on longer tube runs at least 0.5m of smooth bore flexible tubing is

connected to the inlet and discharge port of the pumphead to help to minimise
impulse losses and pulsation in the pipeline. This is especially important with viscous
fluids and when connecting to rigid pipework.

Do

site the pump at or just below the level of the fluid to be pumped if possible.

This will ensure flooded suction and maximum pumping efficiency.

Do

keep the pumphead track and all moving parts clean and free from contamina-

tion and debris.

Do

run at slow speed when pumping viscous fluids. Flooded suction will enhance

pumping performance in all cases, particularly for materials of a viscous nature.

Do

recalibrate after changing pump tubes, fluid, or any connecting pipework. It is

also recommended that the pump is recalibrated periodically to maintain accuracy.

Tube selection

: The chemical compatibility lists published in Watson-Marlow pub-

lications are guides. If in doubt about the compatibility of a tube material and the
duty fluid, request a Watson-Marlow tube sample card for immersion trials.

4.3 Troubleshooting

If the pump runs but there is little or no flow, make the following checks:

Check that the tube is in the pumphead.

Check that fluid is supplied to the pump.

Check that the tube is not split or burst.

Check for any kinks or blockages in the lines.

Check that any valves in the lines are open.

Check that the correct wall-thickness tube is being used.

Check direction of rotation.

114DV, 114DVP

114DV, 114DVP

Содержание Alitea 114 Series

Страница 1: ...fications 4 3 1 Dimensions 4 4 Good pump installation practice 5 4 1 General recommendations 5 4 2 Do s and do not s 6 4 3 Troubleshooting 6 5 Pumphead mounting 7 6 Torque requirement 8 7 Tube holder...

Страница 2: ...Fabriksvag 29 31 120 33 Stockholm Sweden Telephone 46 8 556 556 00 Fax 46 8 556 556 25 The information in this user guide is believed to be correct at the time of publica tion However Watson Marlow Al...

Страница 3: ...Work Act 1974 There are moving parts inside the pumphead Before opening the flip top ensure that the following safety directions are followed Ensure that the pump is isolated from the mains power Ensu...

Страница 4: ...tinuous flow and up to 510 ml min intermittent flow Two tube holder positions to accept tube in bore sizes from 0 5mm to 4 8mm Models with occlusion settings for standard or high pressure operation Un...

Страница 5: ...r if the rotor rotates clockwise and that performance against pressure will be maximised if the rotor rotates counter clockwise Peristaltic pumps are self priming and self sealing against backflow No...

Страница 6: ...lse losses and pulsation in the pipeline This is especially important with viscous fluids and when connecting to rigid pipework Do site the pump at or just below the level of the fluid to be pumped if...

Страница 7: ...Mount the pumphead on its supporting plate through the holes indicated below using two Allen key headed M4 screws length equal to 5mm the thickness of the mount Tighten them to a maximum of 0 5Nm Driv...

Страница 8: ...he above figures x 2 Assess your application s torque requirement from the top of the arrows if all the parameters listed are those requiring more torque as shown in the graph below or to be sure to h...

Страница 9: ...and wandering across the rollers causing premature tube rupture With the larger bore tubes of 4 0mm and 4 8mm the outer position must be used to prevent the flow rate being excessively reduced For tub...

Страница 10: ...to its correct alignment If it does not rise repeat the procedure being sure to maintain downward pressure until release Adjust the tube holder on the other side of the pumphead in the same way To cha...

Страница 11: ...ck pressed against the pumphead inner wall The tube must not be twisted or stretched against the rollers Lower the flip top until it clicks into its fully closed position The track closes automaticall...

Страница 12: ...14 29 47 5 67 85 190 rpm 4 3 8 2 26 5 55 90 5 128 160 200 rpm 4 6 8 6 28 58 95 135 170 350 rpm 8 0 15 49 100 165 235 300 400 rpm 9 1 17 56 115 190 270 340 600 rpm 13 5 26 84 175 285 405 510 Flow rate...

Страница 13: ...e rotation optimizes the pump performance regarding tube life and required torque Clockwise rotation is recommended for low or no pressure applica tions Suction performance and flow are the same as fo...

Страница 14: ...2 0024 016 933 0024 016 913 A024 016 3 2 16 902 0032 016 933 0032 016 913 A032 016 4 0 933 0040 016 4 8 25 902 0048 016 933 0048 016 913 A048 016 mm Tygon E3603 Neoprene 0 8 13 920 0008 016 1 6 14 E36...

Страница 15: ...aint Gobain Corporation STA PURE PFL and STA PURE PCS are trademarks of W L Gore and Associates 12 Warning not to use pumps in patient connected applications Warning These products are not designed fo...

Отзывы: