![Watson Marlow Pumps 530 Series Скачать руководство пользователя страница 33](http://html2.mh-extra.com/html/watson-marlow-pumps/530-series/530-series_operating-manual_3763085033.webp)
18
MemoDose mode
Each time the pump is started by pressing
START
, it records the number of pumphead revolutions which
occur until
STOP
is pressed. The number of revolutions is proportional to the volume of fluid which has been
dispensed: the dose. MemoDose mode allows the user to repeat-dose a precise volume of fluid. This can be
achieved by dispensing a master dose, or entering a dose volume manually using the keypad. MemoDose
can repeat this dose exactly or proportionately.
18.1
To configure MemoDose
o
Select
MODE
o
Using the
∧
/
∨
keys, scroll to
MemoDose
and press
SETTINGS
Note: the pump must be stopped to enter MemoDose Settings.
m-530sn-gb-02
33
Содержание 530 Series
Страница 3: ...27 Trademarks 85 28 Disclaimers 86 29 Publication history 87 m 530sn gb 02 3...
Страница 14: ...14 m 530sn gb 02 8 3 Pumphead options 530 pump range 520R 501RL 313 314 505L 505BA 505CA 314MC 318MC...
Страница 87: ...29 Publication history m 530sn gb 02 530 S SN pump First published 12 15 Issue 2 06 16 m 530sn gb 02 87...