Operating in APPI mode
3-7
APCI probe position
5. Set the parameters to the recommended values shown in the following
table.
Recommended Tune window APPI+ Source tab parameters
Parameter
Recommended value
Voltages
Repeller (kV)
1.50
Cone (V)
55
Extractor (V)
2
RF Lens (V)
0.0
Temperatures
Source Temp (°C)
120
APcI/APPI Probe Temp (°C)
500
Gas Flow
Desolvation (L/hr)
50.0
Cone (L/hr)
150.0
APCI probe
Sample cone tip
Repeller
electrode
Lamp window