background image

Column heater

Sample manager

Binary solvent manager

The ACQUITY UPLC I-Class Series core system includes a binary solvent manager, a sample
manager - flow through needle or sample manager - fixed loop (SM-FTN or SM-FL), a column
heater (CH-A), detectors (tunable ultraviolet, photodiode array, eλ photodiode array, or mass
spectrometry), and an ACQUITY UPLC column.

Waters Empower chromatography data software, MassLynx mass spectrometry software, stand-
alone ACQUITY UPLC Console software, or Analyst software controls the system.

1.7.1 Binary solvent manager (BSM)

The Waters ACQUITY UPLC I-Class Series binary solvent manager (BSM) delivers solvent
compositions for isocratic and binary gradient methods at flow rates of 0.01 to 2.0 mL per minute.
Its features include in-line filters upstream of a primary check valve, the Waters Intelligent Intake
Valve (i2Valve), automated priming functions, and daily system-setup routines.

The design of the binary solvent manager is optimized for sub-2-μm particle liquid
chromatography. The upper pressure limit is set at 124,106 kPa (1241 bar, 18,000 psi) per pump.
A maximum flow rate of 1 mL/min is permitted at system backpressures to 124,106 kPa (1241
bar, 18,000 psi), which falls within the range of optimal linear velocity for sub-2-μm particle
columns of ID 1.0 mm to 3.0 mm.

See also: 

ACQUITY UPLC Binary Solvent Manager PLUS Overview and Maintenance Guide

 on

your documentation media, or at 

Waters.com

, for more details.

Note: 

An optional degasser vent line extension exists for applications requiring longer tubing

lengths. For more information, contact your local Waters distributor.

1.7.2 Sample manager-FTN (SM-FTN)

The ACQUITY UPLC I-Class Series sample manager - flow through needle (SM-FTN) differs
from a loop-based injector. The mechanism aspirates a sample from plates and vials and holds it
in the sample needle in preparation for injecting the sample onto the column. The needle serves
as part of the injection flow path when the sample is pushed onto the column.

Optional extension loops (installed between the sample needle and the injection valve) can
increase the injection volume beyond that of the sample needle. The SM-FTN can also dilute
samples using the auto-dilution option.

Using the flow through needle (FTN) mechanism, the system operates similarly to most traditional
HPLC systems, which facilitate the transfer of HPLC methods. The FTN mechanism also does
not require you to learn new injection modes. Also, it achieves injection accuracy and decreases

April 4, 2018, 715005704 Rev. A

Page 18

Содержание ACQUITY UPLC I-Class Series

Страница 1: ...ACQUITY UPLC I Class Series System Guide 715005704 Revision A Copyright Waters Corporation 2018 All rights reserved...

Страница 2: ...Waters website waters com Trademarks ACQUITY is a registered trademark of Waters Corporation ACQUITY UPLC is a registered trademark of Waters Corporation Analyst is a registered trademark of Applied B...

Страница 3: ...ters Corporation XBridge is a registered trademark of Waters Corporation Waters is a registered trademark of Waters Corporation All other trademarks are property of their respective owners Customer co...

Страница 4: ...tentially hazardous effects of all substances you work with Always follow Good Laboratory Practice GLP and consult your organization s standard operating procedures as well as your local requirements...

Страница 5: ...void injury from electrical shock or fire and damage to the equipment follow these guidelines Do not expose the workstation or ancillary equipment to dripping or splashing liquids Do not place objects...

Страница 6: ...quality control QC procedures and the guidelines presented in this section Applicable symbols The following symbols can be present on the device system or packaging Symbol Definition Manufacturer Date...

Страница 7: ...he system is for research use only and is not intended for use in diagnostic applications Calibrating To calibrate LC systems adopt acceptable calibration methods using at least five standards to gene...

Страница 8: ...irements Group 1 products apply to intentionally generated and or used conductively coupled radio frequency energy that is necessary for the internal functioning of the equipment Class B products are...

Страница 9: ...afety notice v Equipment misuse notice vi Safety advisories vi Operating the system vi Applicable symbols vi Audience and purpose vii Intended use of the ACQUITY UPLC I Class Series system vii Calibra...

Страница 10: ...lexCart optional 21 1 7 11 Column technology 21 1 7 12 Detection 22 1 8 For additional information 22 2 Performance optimization 24 2 1 General guidelines 24 2 2 Dispersion 25 2 3 Carryover 26 2 4 Con...

Страница 11: ...2 Sample manager control panel 50 3 5 3 TUV control panel 51 3 5 4 Column manager control panel 52 A External connections 54 A 1 Ethernet connections 54 A 2 Column heater connection 54 A 3 External c...

Страница 12: ...ents and devices 71 B 4 Electrical and handling symbols 76 B 4 1 Electrical symbols 76 B 4 2 Handling symbols 76 B 5 Stacking system modules with interlocking features 77 B 6 Stacking system modules w...

Страница 13: ...ntation and column design to introduce UPLC technology to the field of separation science By employing this technology Waters ACQUITY UPLC systems achieve a marked increase in resolution speed and sen...

Страница 14: ...ure above that using 1 7 m particles achieves higher efficiency that persists as flow rate increases lower HETP indicates higher efficiency When operating in this area of the plot the peak capacity an...

Страница 15: ...conditions for the separations were identical except for the flow rate which was scaled based on particle size 1 2 Features of the ACQUITY UPLC I Class Series system The ACQUITY UPLC I Class Series s...

Страница 16: ...ds to other systems UPLC applications can benefit from pre column mobile phase heating to improve chromatographic separations The ACQUITY UPLC I Class Series column compartments use an active preheate...

Страница 17: ...ation You also benefit from improvements in sensitivity and flexibility See also Waters com for more information on compatible Waters mass spectrometers 1 7 System components The following illustratio...

Страница 18: ...6 kPa 1241 bar 18 000 psi which falls within the range of optimal linear velocity for sub 2 m particle columns of ID 1 0 mm to 3 0 mm See also ACQUITY UPLC Binary Solvent Manager PLUS Overview and Mai...

Страница 19: ...ature variations can shift peak retention times and alter peak shapes increasing the difficulty of achieving precise results The ACQUITY UPLC I Class Series system s column heater CH A helps to ensure...

Страница 20: ...to 65 0 C The 30 cm CHC module shares many of the same components with the 30 cm CH module except for a thermoelectric engine for heating and cooling and an additional fan for exhausting engine heat I...

Страница 21: ...ides electrical outlets for system instruments and integrated waste management Used with a mass spectrometer the cart s adjustable height lets you position the column outlet close to the inlet probe m...

Страница 22: ...ificate of analysis QC test data When you attach the column s eCord fob to the receptacle on the column compartment the chip automatically records and stores system information You need take no furthe...

Страница 23: ...find more information and to join the ACQUITY UPLC online community where you can Share information and ask questions of ACQUITY UPLC experts and scientists Access ACQUITY UPLC publications and user...

Страница 24: ...nalyses a peak of interest can be less than 0 5 seconds Waters recommends a sampling rate of 25 to 50 points across the peak which provides good quantitation and peak representation Sampling rates fas...

Страница 25: ...on UPLC systems exhibit low dispersion a fixed instrument characteristic measured by the extent of peak broadening which occurs because of the system volume Small particle chromatography uses small hi...

Страница 26: ...ad Ahead and Loop Offline options Initiating these options before the highly organic part of the gradient reaches the needle can leave sample residue in the system The time savings you gain from these...

Страница 27: ...specifically that different techniques apply to retightening fittings versus installing them for the first time See also The user documentation for the modules included in your system for more details...

Страница 28: ...or set to tubing you must mark the compression screw and connection port and ensure that the two marks line up when you tighten them Warning To avoid eye injury use eye protection when performing this...

Страница 29: ...se 3 4 turn until the two marks line up First use tightening 2 7 1 2 Stainless steel gold plated fitting with long flats and 2 piece stainless steel ferrule V detail First use Long flats Compression s...

Страница 30: ...led Long flats Compression screw 2 piece stainless steel ferrule Tighten the fitting finger tight plus as much as an additional 1 6 turn using a 1 4 inch open end wrench Reinstalled tightening April 4...

Страница 31: ...ghten the fitting finger tight plus an additional 3 4 turn using a 1 4 inch open end wrench For detailed instructions about assembling new fittings see Assembling new fittings Tip To prevent band spre...

Страница 32: ...g finger tight plus as much as an additional 1 6 turn using a 1 4 inch open end wrench Reinstalled tightening 2 7 1 4 1 4 28 flangeless fitting with ferrule First use or re installed 1 Compression scr...

Страница 33: ...en the fitting finger tight 2 7 1 6 Long 1 4 28 fitting with flangeless ferrule and stainless steel lock ring installed on 1 8 inch OD tubing First use or re installed Compression screw Lock ring Ferr...

Страница 34: ...w Lock ring Ferrule End of lock ring with larger inside diameter ID Tighten the fitting finger tight 2 7 1 8 5 16 24 fitting with filter and stainless steel lock ring First use or re installed Compres...

Страница 35: ...rst use or reinstalled Compression screw Tighten the fitting finger tight Tips You can also use the column gripping tool when tightening this fitting To prevent band spreading ensure that the tubing i...

Страница 36: ...ed 1 knurled compression fitting 2 locking cap nut Back locking PEEK ferrule Figure 2 3 Column stabilizer outlet fitting first use or reinstalled 1 hex compression fitting 2 locking cap nut Back locki...

Страница 37: ...e in line filter Important To prevent band spreading ensure that the tubing bottoms in the connection port before you tighten the compression fitting 3 Finger tighten the compression fitting into the...

Страница 38: ...et ferrule on the APH assembly fitting if the column is leaking or the ferrule looks damaged Warning To prevent burn injuries set the column temperature to Off and then allow the column compartment an...

Страница 39: ...mpression fitting off the tubing Figure 2 7 Sliding fitting off tubing 1 knurled compression fitting Tubing Note If the assembly contains a captive ferrule the ferrule remains locked in the fitting 3...

Страница 40: ...ule on the fitting 6 Slide the 1 knurled compression fitting and ferrule onto the tubing Figure 2 9 Sliding fitting and ferrule onto tubing Tubing 1 knurled compression fitting Note If it is a captive...

Страница 41: ...ons on choosing and handling solvents 2 8 2 Matching sample diluents When you use the auto dilution option on the SM FTN the purge solvent serves as the sample diluent Ensure that your sample solution...

Страница 42: ...g start up tests run CPU board memory RAM and ROM external communication system Ethernet and clock If the start up tests indicate a malfunction contact your local Waters representative 2 Power on the...

Страница 43: ...select Run Samples Result A control panel for each device in the system appears 2 In the sample manager control panel click Display console Alternative Right click the control panel for any module and...

Страница 44: ...he following start up tests run CPU board memory RAM and ROM external communication system Ethernet and clock If the start up tests indicate a malfunction contact your local Waters representative 2 Po...

Страница 45: ...ly even when the power switch is in the off position These fans switch off only when you disconnect the power cable from the ac wall outlet or rear panel 3 3 4 Status LEDs 3 3 4 1 Flow LED Flow status...

Страница 46: ...r is operating normally completing any outstanding samples or diagnostic function requests Flashing green The sample manager is initializing Flashing red An error stopped the sample manager Refer to t...

Страница 47: ...tus unless previously enabled To enable the leak sensors 1 In the ACQUITY UPLC Console select Control Leak Sensors 2 To enable the leak sensor for an individual module click the status on the left han...

Страница 48: ...e all solvents and to specify the solvent composition flow rate column and sample temperatures and needle characterization for your next system start up For details see the console online Help 3 5 Mon...

Страница 49: ...values range from 0 0 to 100 0 Stop flow Immediately stops all flow from the binary solvent manager Pressure Delta The difference between the maximum and minimum pressures observed during the previous...

Страница 50: ...can edit these values when the system is idle by clicking the underlined value You cannot edit sample manager set points while the system is running samples Tip To keep the sample compartment at a co...

Страница 51: ...e sample manager control panel Control panel function Description Prime syringes Displays the Prime dialog box Wash needle Displays the Wash Needle dialog box Turn lights on Illuminates the ACQUITY UP...

Страница 52: ...anel Table 3 6 Additional functions in the TUV detector control panel Control panel function Description Auto zero Resets the absorbance value to 0 Reset TUV Resets the detector after an error conditi...

Страница 53: ...ys the current column compartment temperature Temperature set point Displays the column compartment set point When active temperature control is disabled this field displays Off You can access these a...

Страница 54: ...em modules Connect the shielded Ethernet cables from each module to the electronic connections on the rear panel of the sample manager The sample manager is connected internally to the Ethernet switch...

Страница 55: ...ed off 2 Connect the external communication cable to the High Density HD port on the rear of the column heater 3 Connect the other end of the external communication cable to the PSPI port on the rear...

Страница 56: ...A 3 External cable connections Figure A 1 System rear panel cable connections April 4 2018 715005704 Rev A Page 56...

Страница 57: ...r Sample manager Solvent manager Sample organizer Ethernet switch Workstation Interconnect D sub cables Ethernet cables A 4 Plumbing connections Figure A 2 System plumbing connections April 4 2018 715...

Страница 58: ...Detector Column heater Sample manager Solvent manager Sample organizer Stainless steel tubing PEEK tubing April 4 2018 715005704 Rev A Page 58...

Страница 59: ...A 5 Waste tubing connections Figure A 3 System waste tubing connections April 4 2018 715005704 Rev A Page 59...

Страница 60: ...erforming any maintenance operation on it Connect each module to a common ground Note Some column modules such as the column heater CH A the column heater 30 cm CH 30A and the column manager auxiliary...

Страница 61: ...driver Flat blade screwdriver Connector Signal cable To connect the cables 1 Insert the connector into the connector port on the module s rear panel Figure A 4 Inserting connector into connector port...

Страница 62: ...minal on the rear panel grounding stud and secure the terminal using the locking nut Note Use the 9 32 inch nut driver to tighten the locking nut until the terminal does not move Figure A 6 Grounding...

Страница 63: ...r I O signal cables These connectors are keyed so that they can be inserted only one way Figure A 7 Binary solvent manager I O signal connectors Connector 1 Auxiliary 1 In Auxiliary 1 In Auxiliary 2 I...

Страница 64: ...2V Analog 2 Out 0 2V Analog 2 Out For electrical specifications see the Binary Solvent Manager Overview and Maintenance Guide on your documentation media or at Waters com for more details Table A 1 Bi...

Страница 65: ...ware failure occurs on another system module Switch 2 Out Used to send time based contact closure signals to external devices Switch 3 Out Used to send time based contact closure signals to external d...

Страница 66: ...Inject Start Out Indicates with a contact closure output that an injection has started Inject Hold In Delays the next injection when the sample manager receives a contact closure input from another sy...

Страница 67: ...er supplies to operate To connect the column module Notice To avoid damaging the electronic components of the sample manager and the column heater or column heater cooler always power off the sample m...

Страница 68: ...liability in cases of injury or property damage resulting from the failure of individuals to comply with any safety precaution when installing repairing or operating any of its instruments or devices...

Страница 69: ...gical agents that can pose a serious health threat Warning Risk of tipping Warning Risk of explosion Warning Risk of high pressure gas release B 1 1 Specific warnings The following warnings both symbo...

Страница 70: ...0 psi during an analysis requiring the use of flammable solvents Also a gas fail device must be installed to interrupt the flow of LC solvent should the nitrogen supply fail B 1 1 3 Mass spectrometer...

Страница 71: ...st edition of the National Research Council s publication Prudent Practices in the Laboratory Handling and Management of Chemical Hazards Observe Good Laboratory Practice GLP at all times particularly...

Страница 72: ...iritto all utilizzo dell apparecchiatura da parte dell utente Advertencia cualquier cambio o modificaci n efectuado en esta unidad que no haya sido expresamente aprobado por la parte responsable del c...

Страница 73: ...lle offenen Flammen in der N he l schen Keine Schl uche verwenden die stark geknickt oder berbeansprucht sind Nichtmetallische Schl uche nicht f r Tetrahydrofuran THF oder konzentrierte Salpeter oder...

Страница 74: ...trabajar con tubos que se hayan doblado o sometido a altas presiones Es necesario utilizar tubos de metal cuando se trabaje con tetrahidrofurano THF o cidos n trico o sulf rico concentrados Hay que t...

Страница 75: ...tion assur e par le mat riel risque d tre d fectueuses Warnung Der Benutzer wird darauf aufmerksam gemacht dass bei unsachgem er Verwendung des Ger tes die eingebauten Sicherheitseinrichtungen unter U...

Страница 76: ...Electrical power off Standby Direct current Alternating current Alternating current 3 phase Safety ground Frame or chassis terminal connection Fuse Functional ground Input Output B 4 2 Handling symbo...

Страница 77: ...with interlocking features Warning To avoid spinal and muscular injury do not attempt to lift a system module without assistance Warning To avoid crushing your fingers beneath or between modules use...

Страница 78: ...e remaining system modules B 6 Stacking system modules without interlocking features This procedure applies to system modules that are not equipped with interlocking features Most modules will however...

Страница 79: ...e 2 Carefully lower the module so that the feet rest in the indents Important To maintain the integrity of the system stack and integrated waste system ensure that the feet of the upper module rest in...

Отзывы: