2.5.1.10 One-piece PEEK fitting
Figure 2–1: First use or reinstalled
Compression screw
Tighten the fitting finger-tight.
Tips:
• You can also use the column gripping tool when tightening this fitting.
• To prevent band spreading, ensure that the tubing is fully bottomed in the connection port
before tightening the fitting.
2.5.1.11 Dual-threaded fitting with locking cap nut
Required tools and materials
• Column gripping tool
Figure 2–2: APH outlet fitting first use or reinstalled
#1 knurled compression fitting
#2 locking cap nut
Back-locking PEEK ferrule
April 4, 2018, 715005702 Rev. A
Page 37