Replacing syringe plungers and plunger tips
3-15
4.
Remove the Teflon syringe gasket from the valve’s syringe fitting port.
5.
Fill the new syringe with degassed wash solvent.
Recommendations:
•
Use isopropanol as the wash solvent.
•
Remove air bubbles from the syringe.
6.
Fit a new Teflon gasket on the syringe.
Tip:
The syringe includes a Teflon syringe gasket.
7.
Connect the plunger of the filled syringe to the syringe drive by pushing
it straight onto the clip.
Requirement:
Keep the top threaded end below the syringe valve.
8.
Raise the syringe, and screw it firmly into the connector.
9.
In the Replace Sample Syringe window, click Reset.
Result:
The syringe moves to the home position and dispenses its
contents to waste.
10. Using the Control menu of the 2707 Console, perform a standard wash
routine.
Result:
All tubing connected to the syringe valve refills and flushes.
11. Purge all air bubbles from the syringe and buffer tubing by repeatedly
washing the needle or removing and repriming the syringe with
degassed, nonaqueous organic solvent.
Replacing syringe plungers and plunger tips
To replace the plunger or plunger tip
1.
From the Maintain menu of the 2707 Console, click Replace > Syringe.
Result:
The syringe plunger lowers.
2.
Unscrew the syringe from syringe valve.
Requirement:
Keep the connector in the valve intact.
Caution:
To avoid valve damage, do not use a sharp tool for gasket
removal.