Exploring injection principles
1-23
6.
The injection valve switches to the Inject position, and the sample loop
becomes part of the flow path for the HPLC mobile phase. Sample enters
the column, and analysis starts.
After each injection, a wash routine of programmable volume occurs.
Wash station
Buffer tubing
Syringe
Needle
Sample loop
INJECT
Column
Pump
Sample