Section 5 – Probe Calibration
23
Section 5. – Probe Calibration
Your
Chemflo
is supplied pre-calibrated. However the calibration of the pH probe should be
checked periodically.
5.1 pH Calibration
5.1.1 pH Calibration Procedure
1. Press .
2. Press until “
o
Dosing Mode
” is selected and press
.
The pH Calibration screen is now displayed. For example…
7.19pH Read&Set
7.19pH Enter
3. Obtain a value of the pool water pH using a pool test kit or have the water tested.
4. Press or
to input the pH value.
5. Press when the pH reading on the top line becomes stable.
5.1.2 pH Calibration Notes
1. The
Chemflo
pH probe has an allowable Asymmetry range of –1.00 to +1.00 pH. If calibration
fails due to the Asymmetry being outside these limits, please consult the Troubleshooting
guide (section 7.4) for possible remedies.
2. All calibration information is retained in memory when the
Chemflo
is switched off, even when
the power supply is removed.
5.1.3 pH Calibration Messages
1. If a calibration has been successfully performed, the
Chemflo
will display the following
message and the asymmetry of the probe. For example…
Calibrate OK
Asym=0.10pH
2. If a calibration has failed, the
Chemflo
will display the following message and the failed
asymmetry value of the probe. For example…
Calibrate Failed
Asym=1.10pH