CLP
OPERATOR’S MANUAL
#96-6294 • Revised 7/04
PH402 CALIBRATION PROCEDURES
Press the MODE key, AUT.CAL appears in the display, and the
YES/NO key prompt flags flash. Press YES.
NEW.SNS: Display flashes YES/NO
answer YES if new sensor fitted or NO if not.
(Take Care! Answering YES resets logbook calibration data.)
Insert the sensors in pH 7 buffer solution.
Press YES to start calibration.
Note: To start calibration with another solution,
press NO to cycle through the other options.
→ → → → →
NO
Rinse sensor.
Transfer to second buffer (pH4) and press YES.
The display now flashes the pH
value. The instrument waits for
the reading to stabilize.
When the reading is stable, the CAL END
message appears. Press Yes.
After briefly displaying WAIT, the display
returns to the normal readout
The instruments wait for
the reading to stabilize.
(The pH display flashes.)
When reading is stable,
the CAL END message
appears.
↓↓↓↓↓
YES
→
→
→
→
→
NO
→
→
→
→
→
YES