![Water Maze CLP Series Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/water-maze/clp-series/clp-series_operators-manual_3560895031.webp)
CLP
OPERATOR’S MANUAL
31
#96-6294 • Revised 7/04
DIGITAL ORP/PH
CONTROLLER - MODEL PH402
Explanation of Operating Keys
MODE Key
This key toggles between the measuring
and maintenance modes. Press once to
obtain access to the maintenance
function menu.
AUTO CAL
MAN CAL
DISPLAY
SETPOINT
WASH
MAN.IMP
TEMPERATURE
HOLD
Press again to return to the measuring
mode (press twice when hold is activated.
YES/NO Keys These are used to select choices from
the menu.
YES is used to accept a menu selection.
NO is used to reject a selection or to
move ahead to the next option.
DATA ENTRY Keys (> ^ ENT )
>
is used as a "cursor" key. Each press
on this key moves the cursor or flash-
ing digit one place to the right.
^
is used to change the value of a se-
lected digit. Each press on this key in-
creases the value by one unit. The
value can not be decreased, so in or-
der to obtain a lower value, increase
past 9 to zero, then increase to the
required number.
ENT
When the required value has been set
using the > & ^ keys, press ENT to
confirm the data entry. Please note
that the controller does not register
any change of data until the ENT key
is pressed.
* Key
This is the commissioning mode key.
It is used to obtain access to the com-
missioning menu. This can only be
done with the cover removed or
opened. Once this button has been
used to initiate the commissioning
menu, follow the prompts and use the
other keys as described above.
PH402 OPERATOR INTERFACE