Warnings
Read Before Operating
The AirWatch by WatchGas is a multigas detector that
can turn machines on or off. It is suitable for almost
every application. The AirWatch, which can be used as
a safe area monitor.
The basic use of the device is the monitoring of ma-
chines in safe zones and turning them off if gasses are
detected
.
The AirWatch MUST NOT be used in a
potentially explosive atmosphere!
Check whether the AirWatch has a calibration sticker
showing when it needs to be checked and recalibra-
ted. We recommend having it checked and calibrated
every six months. Test prior to use!
In order to turn the AirWatch on, it must always be
provided with a power source using the 5-pin Amp-
henol connection (cable type 1) or a 3-pin with battery
clamps (cable type 2).
Never open the device when its running. Shut
off power and de-energize 15 minutes before
opening the AirWatch.