Instruction Manual MODEL AM-756

 

9/20

 

5-3 Displaying Scaling Settings 

(1)

 

Set the full scale display value to "10000". 

Mode

3 times

Mode

Mode

/

Change the set value 
using the above keys

,

Mode

 

 
(2)

 

Set the full scale input value to "5.000". 

Mode

Mode

/

Change the set value 
using the above keys

,

 

 
(3)

 

Change the offset display value to "0". (As the default value is "0", this will not be changed.) 

Mode

Check value is correct

Enter

 

 
(4)

 

Change the offset input value to "1.000". 

Mode

Mode

/

Change the set value 
using the above keys

,

 

 
(5)

 

Set the position of the decimal point to "0.00". 

Mode

Mode

3 times

Enter

Enter

 

 

5-4 Comparative Output Settings 

(1)

 

Set the Comparative Output HI Comparative Value to "60.00". 

(As the setting condition is HH > HI, the value of HH cannot be lower than the value of HI. Therefore, enter the HI value first.) 

Mode

3 times

Mode

/

,

Change the set value 
using the above keys

Mode

Mode

 

 
(2)

 

Set the comparative output HH comparative value to "80.00". 

Mode

/

/

Change the set value 
using the above keys

,

Mode

 

 
(3)

 

Set the comparative output LO comparative value to "30.00". 

Mode

/

Change the set value 
using the above keys

,

Mode

Twice

 

 
(4)

 

Set the comparative output LL value to "10.00". 

Mode

/

Change the set value 
using the above keys

,

Mode

 

 

5-5 Return to Measurement Mode 

Return to measurement mode and commence measurement. 

Enter

Enter

Enter

Measurement mode

 

 

Содержание AM-756

Страница 1: ...s 1 Watanabe Electric Industry Co Ltd assumes no responsibility for any special damage indirect loss or passive loss caused by this device 2 This device has been developed designed and manufactured on...

Страница 2: ...elow Min 120mm Min 70mm 0 6 0 45 92 0 0 8 2 2 External Dimensions 48 96 91 5 102 5 77 86 5 97 5 21 4 11 0 5 2 Mode Enter 88 5 2 3 Mounting the Unit on the Panel 1 After removing the mounting bands fro...

Страница 3: ...No Name Details 1 NC Unconnected Please do not use these terminals as relay terminals 2 NC 3 NC 4 HI Terminal 5 LO Terminals connected internally 6 LO Input range settings cannot be made 3 3 3 DB Ran...

Страница 4: ...nal 5 GO C GO Output photocoupler Collector terminal 6 GO E GO Output photocoupler Emitter terminal 7 LO C LO Output photocoupler Collector terminal 8 LO E LO Output photocoupler Emitter terminal 9 LL...

Страница 5: ...ut Display Displays comparative judgment results Extinguished Comparative output OFF 2 Main Display Displays measurement values Displays group name parameters set values 3 Function Indicators RE Unuse...

Страница 6: ...ode Enter Mode Enter Mode or Mode Enter Mode or Mode Mode Enter Mode or Mode Enter Mode or Mode Enter Mode or Mode Enter Mode Enter Mode or Mode Enter Mode or Mode Enter Mode Enter Mode or Mode Enter...

Страница 7: ...Process value Input unit Comparative output unit Control terminals 5 7 Key repeat will occur when the or keys are held down 4 5 Changing Parameter Settings The diagram below shows the basic operating...

Страница 8: ...thin the range or cancel the setting change Canceling Changes to Set Values Pressing the Enter key during the changing of a set value cancels the value change The set value will revert to the value pr...

Страница 9: ...e 3 times Enter Enter 5 4 Comparative Output Settings 1 Set the Comparative Output HI Comparative Value to 60 00 As the setting condition is HH HI the value of HH cannot be lower than the value of HI...

Страница 10: ...zero control method Option any apart from NO Digital Zero Value 199999 to 199999 digit 0 Sets the digital zero value when digital zero is ON N B If settings are made while digital zero is OFF they wi...

Страница 11: ...o 199999 0 Sets the display value when input is offset display values to Offset Input Values 1 to 8 Changes according to input range See chart below See chart below Sets the input value for offset dis...

Страница 12: ...t value LO output ON Decimal point position depends on the decimal point position in identical patterns to LL Comparison Set Values 1 to 8 100000 Measurement value LL set value LL output ON Decimal po...

Страница 13: ...etting Condition N10i N09i N08i 8000 Input Value 9 Linearize pre correction display Output Value 9 Linearize post correction display LINE Set Value 16 Linearize Output Value 9 199999 to 199999 digit S...

Страница 14: ...t Value Input Range Possible Display Range Resolution 11 0 2V Approx 0 21V 199999 12 2V Approx 2 1V 13 20V Approx 21V 14 200V Approx 210V 23 20mA Approx 21mA 24 200mA Approx 210mA 1V 5V Approx 5 25V 2...

Страница 15: ...is activated when Tracking zero correction width Measurement value Tracking zero correction width Parameter Used 7 3 3 Step Wide Function This function changes the resolution of the least significant...

Страница 16: ...ontrol terminal 2 Operation Type There are two Start Hold operation types Type A free run and Type B one shot Parameter Used Type A This operation type enters hold status when Start Hold is ON halting...

Страница 17: ...h output are undefined at start up N B 2 Open collector output is NPN open collector output N B 3 When the computed result is WAIT the data output will be 000000 2 Over range Signal OVER Output is act...

Страница 18: ...power External dimensions 96mm W 48mm H 97 5mm D without optional unit installed Weight Approx 200 to 300g Insulation withstands Voltage AC 500V per minute between input terminal and each of the outp...

Страница 19: ...the main unit connector female 9 2 Trouble Shooting Problem Possible Cause Solution The monitor display is blank The display blank function is set to extinguish display Turn OFF the extinguish display...

Страница 20: ...e Electric Industry Co Ltd Watanabe Electric Industry Co Ltd will make repairs free of charge 11 2 After sales Service This product was manufactured tested inspected and shipped under strict quality c...

Отзывы: