Instruction Manual MODEL AM-756

 

3/20

 

Terminal Description and Connection Method 

3-1 Rear View 

Input unit

Comparative 

output unit

Optional unit

Power unit

 

 
 
 

 

 

3-2 Power Connection 

M3 screw terminal (terminal cover included) 

NC

Top

Bottom

AC power drive

100 to 240VAC

±10%

12 to 48VDC

±10%

NC

NC

NC

1

2

3

4

5

6

NC

NC

DC power drive

NC

NC

1

2

3

4

5

6

Applicable crimp 

terminal

Max. 5.8

Max. 5.8

Top

Bottom

Recommended

torque : 0.5Nm

 

*NC Terminal: unconnected 
 
 
 

 

 

3-3 Input Connection 

M3 screw terminal (terminal cover included) 

3-3-1 DV Range (Direct voltage input) 

11 HI

11 LO

14 HI

12,13,14 LO

13 HI

12 HI

1

2

3

4

5

6

Max. 5.8

Max. 5.8

Top

Bottom

Applicable crimp 

terminal

Recommended

torque : 0.5Nm

 

 

Terminal 

No.

 

Name 

Details 

14 HI 

±199.99V input (14 range) + terminal 

 

12,13,14 LO 

12, 13, 14 range - terminals 

13 HI 

±19.9999V input (13 range) + terminal 

12 HI 

±1.99999V input (12 range) + terminal 

11 HI 

±0.199999V input (11 range) + terminal 

11 LO 

±0.199999V input (11 range) - terminal 

The input range settings parameters can be used to change the input range. 

* Please do not connect anything to terminals other than those for the range 
used.     

Factory Setting: 11 ranges (± 0.199999V) 

 

3-3-2 23, 24 Range (Direct current input) 

LO

LO

NC

NC

NC

HI

1

2

3

4

5

6

Max. 5.8

Max. 5.8

Top

Bottom

Applicable crimp 

terminal

Recommended

torque : 0.5Nm

 

 

Terminal 

No. 

Name 

Details 

NC 

Unconnected 
* Please do not use these terminals as relay terminals. 

NC 

NC 

HI 

+ Terminal 

LO 

- Terminals (connected internally) 

LO 

* Input range settings cannot be made. 

 

3-3-3 DB Range (Process input) 

Top

Bottom

+EXC

-EXC

2A HI

LO

1V HI

11 HI

1

2

3

4

5

6

Max. 5.8

Max. 5.8

Applicable crimp 

terminal

Recommended

torque : 0.5Nm

 

 

Terminal 

No. 

Name 

Details 

2A HI 

4 to 20mA (±20mA) input (2A range ) + terminal 

LO 

- terminal 

1V HI 

1 to 5 V (±5V) input (1V range) + terminal 

11 HI 

±0.199999V input (11 range) + terminal 

+EXC 

Sensor power  terminal 

-EXC 

Sensor power output - terminal 

The input range settings parameters can be used to change the input range. 

* Please do not connect anything to terminals other than those for the range 
used. 

Factory Setting: 1V range (±5V) 

 

  Switching Sensor Power Voltage 

This device is set to a factory setting of +24V. Therefore, when using the 
device at +12V and +10V, please adjust the sensor power voltage according to 
the instructions below. 
(1)

 

Turn off the device and leave for a minimum period of 10 minutes. Next, 
pushing in the upper and lower hooks on the case, pull out the input unit. 

 

Hooks

(also located on bottom)

Input unit

 

 

(1)

 

Please  ensure  supply  voltage  is  within  device’s  power  rating  and  that 
power-up and cut-off are instantaneous. 

(2)

 

Please wait for ten seconds or more before activating power each time. 

(3)

 

Please do not use the NC terminals as relay terminals. 

(1)

 

Units may not be changed by the customer. 

(2)

 

Do not use different models of these units. 

CAUTION

 

CAUTION

 

Содержание AM-756

Страница 1: ...s 1 Watanabe Electric Industry Co Ltd assumes no responsibility for any special damage indirect loss or passive loss caused by this device 2 This device has been developed designed and manufactured on...

Страница 2: ...elow Min 120mm Min 70mm 0 6 0 45 92 0 0 8 2 2 External Dimensions 48 96 91 5 102 5 77 86 5 97 5 21 4 11 0 5 2 Mode Enter 88 5 2 3 Mounting the Unit on the Panel 1 After removing the mounting bands fro...

Страница 3: ...No Name Details 1 NC Unconnected Please do not use these terminals as relay terminals 2 NC 3 NC 4 HI Terminal 5 LO Terminals connected internally 6 LO Input range settings cannot be made 3 3 3 DB Ran...

Страница 4: ...nal 5 GO C GO Output photocoupler Collector terminal 6 GO E GO Output photocoupler Emitter terminal 7 LO C LO Output photocoupler Collector terminal 8 LO E LO Output photocoupler Emitter terminal 9 LL...

Страница 5: ...ut Display Displays comparative judgment results Extinguished Comparative output OFF 2 Main Display Displays measurement values Displays group name parameters set values 3 Function Indicators RE Unuse...

Страница 6: ...ode Enter Mode Enter Mode or Mode Enter Mode or Mode Mode Enter Mode or Mode Enter Mode or Mode Enter Mode or Mode Enter Mode Enter Mode or Mode Enter Mode or Mode Enter Mode Enter Mode or Mode Enter...

Страница 7: ...Process value Input unit Comparative output unit Control terminals 5 7 Key repeat will occur when the or keys are held down 4 5 Changing Parameter Settings The diagram below shows the basic operating...

Страница 8: ...thin the range or cancel the setting change Canceling Changes to Set Values Pressing the Enter key during the changing of a set value cancels the value change The set value will revert to the value pr...

Страница 9: ...e 3 times Enter Enter 5 4 Comparative Output Settings 1 Set the Comparative Output HI Comparative Value to 60 00 As the setting condition is HH HI the value of HH cannot be lower than the value of HI...

Страница 10: ...zero control method Option any apart from NO Digital Zero Value 199999 to 199999 digit 0 Sets the digital zero value when digital zero is ON N B If settings are made while digital zero is OFF they wi...

Страница 11: ...o 199999 0 Sets the display value when input is offset display values to Offset Input Values 1 to 8 Changes according to input range See chart below See chart below Sets the input value for offset dis...

Страница 12: ...t value LO output ON Decimal point position depends on the decimal point position in identical patterns to LL Comparison Set Values 1 to 8 100000 Measurement value LL set value LL output ON Decimal po...

Страница 13: ...etting Condition N10i N09i N08i 8000 Input Value 9 Linearize pre correction display Output Value 9 Linearize post correction display LINE Set Value 16 Linearize Output Value 9 199999 to 199999 digit S...

Страница 14: ...t Value Input Range Possible Display Range Resolution 11 0 2V Approx 0 21V 199999 12 2V Approx 2 1V 13 20V Approx 21V 14 200V Approx 210V 23 20mA Approx 21mA 24 200mA Approx 210mA 1V 5V Approx 5 25V 2...

Страница 15: ...is activated when Tracking zero correction width Measurement value Tracking zero correction width Parameter Used 7 3 3 Step Wide Function This function changes the resolution of the least significant...

Страница 16: ...ontrol terminal 2 Operation Type There are two Start Hold operation types Type A free run and Type B one shot Parameter Used Type A This operation type enters hold status when Start Hold is ON halting...

Страница 17: ...h output are undefined at start up N B 2 Open collector output is NPN open collector output N B 3 When the computed result is WAIT the data output will be 000000 2 Over range Signal OVER Output is act...

Страница 18: ...power External dimensions 96mm W 48mm H 97 5mm D without optional unit installed Weight Approx 200 to 300g Insulation withstands Voltage AC 500V per minute between input terminal and each of the outp...

Страница 19: ...the main unit connector female 9 2 Trouble Shooting Problem Possible Cause Solution The monitor display is blank The display blank function is set to extinguish display Turn OFF the extinguish display...

Страница 20: ...e Electric Industry Co Ltd Watanabe Electric Industry Co Ltd will make repairs free of charge 11 2 After sales Service This product was manufactured tested inspected and shipped under strict quality c...

Отзывы: