7 Troubleshooting / Correcting errors
Fault
Likely cause
Likely solution
WAPO-Ex does not heat
up
Dirty level sensor
Clean level sensor
Water level too low
Add water
Water-level LED comes
on and warning tone
sounds
Dirty level sensor
Clean level sensor
Water level too low
Add water until red LED
goes out and tone stops
Water pump inoperative
Operating or set
temperature not yet
reached
Heat to operating
temperature (min. 60°C) or
set temperature
Multi-function display not
working correctly
Variations in mains
voltage (220-240 V / 50/60
Hz)
Turn of main switch and
turn on again in about 1
minute.
Multifunction
display shows
A
Water-temperature above
100°C (A3 means 103 °C)
-faulty temperature sensor
-faulty control board
Turn off the water supply,
switch off the unit
completely, contact service
Multifunction
display shows
F 1
Short-circuit:
Faulty electrial component
Turn off the water supply,
switch off the unit
completely, contact service
Multifunction
display shows
F 2
Faulty temperature sensor Turn off the water supply,
switch off the unit
completely, contact service
If the above recommendations do not solve the problem, contact your
dental depot or our Service department.