14
Do’s and Dont’s
Do ensure the following are observed:
The trolley should only be used by properly trained staff at all times.
The trolley must be operated by 2 people when transporting a patient.
When a patient is on board, the cot sides must be in an upright locked position before
moving the trolley or adjusting the trolley height.
The trolley must only be used for the purpose for which it was designed.
Don't:
Lift the trolley upwards by the top.
Push the trolley over rough or uneven ground.
Drop heavy or sharp objects onto the solid plastic table top.
Push the trolley by the cot sides, these are for patient safety only.
Place a weight of over the equivalent of 34 Stone (220Kg/485lbs) onto the trolley.
Place a weight of over the equivalent of 7 Stone 12lbs (45Kg/99lbs) on either of the Pivot
Sections.
Spill oil or other liquids/chemicals on the mattress.
Attempt to rectify any faults on the trolley without contacting Wardray Premise Ltd.
Use the trolley without a mattress.
Do not allow staff or patients to fix or store any items on the trolley unless you have verified
the MR safety of the items.
The trolley should be serviced regularly, but only by Wardray Premise trained
engineers, as they will carry the correct spares to guarantee safety and
compatibility of the trolley. Failure to comply with this will invalidate your warranty
and could compromise its MR compatibility and therefore safety.