31
Disinfection, Cleaning, Sterilization
Motor with cable
Do not twist and kink the motor cable! Do not coil it too tightly!
Manual disinfection
– If heavily soiled: Clean first with disinfectant cloths.
– Disinfect using surface disinfectants; spray disinfection is recommended.
– Use certified surface disinfectants (e.g. DGHM-tested).
For USA:
Use EPA registered surface disinfectants.
– Note the action times of the disinfectants used!
Manual cleaning
– Rinse under demineralized water (< 38 °C) with the aid of a brush (not recommended in U.K.)
– Then remove any liquid residues (absorbent cloth, blow dry with compressed air)
Before disinfection and cleaning
push the motor protective cover
onto the motor attachment!
Made in
AU
S
TRIA