12. Error messages
Error No.
Description
Remedy
00
Electronics overheating –
Switch off the equipment, allow to cool
safety shutdown
for at least 10 minutes, then re-start
01
Electronics overloaded
Switch off the equipment and then re-start
07
Foot control error –
Switch off the equipment, re-start,
do not
initializing
actuate foot control when switching on
09
Foot control error
Switch off the equipment, check connection of
the foot control, then re-start
19
Run limiting control
Switch off the equipment and re-start
99
System failure
Switch off the equipment and re-start
Motor temperature too high
Switch off the equipment, allow motor to cool
for at least 10 minutes, then re-start
Motor connection –
Switch off the equipment, check motor
safety shutdown
connection, then re-start