Diagnostic
108
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
TROUBLESHOOTING TABLE
Notice: Before sending the sterilizer for technical service, remove the mains cable, empty both water tanks and use the original or appropriate
packaging.
Problem
Possible cause
Solutions
The sterilizer remains switched
OFF.
The mains switch or network circuit breaker is OFF.
Activate the mains switch or network circuit breaker (ON).
No voltage at the socket.
Check the electric circuit.
The power cord is not connected properly.
Check and connect the power cord properly.
Water is leaking at the front of the
sterilizer.
Leaks through the chamber door seal.
Clean or replace the door seal.
Clean the chamber face side.
Internal leak.
Call technical service.
The cycle commences but there is
no pressure/temperature rise.
The thermal overload switch is open.
Reset the thermal overload switch. See "Extraordinary maintenance" on page 92.
Electric - electronic fault.
Call technical service.
At the end of the cycle, there is
residual water in the chamber.
Sterilizer not properly leveled.
Properly level the surface the sterilizer is placed on.
Overloaded chamber.
Comply with the maximum load weight limits for each type of load. Always use the
chamber rack for trays and cassettes. See ."Load maintenance and preparation" on
page 54
Chamber filter clogged.
Remove and clean the chamber filter.
Chamber filter cap not well-positioned.
Mount the chamber filter cap properly (see "Ordinary maintenance" on page 73)
Load incorrectly placed.