Diagnostic
102
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
Code
Description and actions
Actions
13x to 16x
Check water level in the clean water tank. Reset the thermal
overload.
Repeat the cycle.
If the problem persists, call service.
Switch the sterilizer OFF and ON.
Clean the door seal and the chamber face side.
Check if the load placed in the sterilization chamber
complies with the MAXIMUM WEIGHT LIMITS.
Clean the chamber and the chamber furniture from
residuals of detergents, disinfectants and other chemicals.
Replace the clean water if it is suspected to be
contaminated with chemicals.
Ensure all the load is clean rinsed and free from any
chemicals before sterilizing.
Start a vacuum test to check the tightness of the
pneumatic circuit.
18x
Chamber filter clogged. Remove and clean the chamber
filter. See error "13x to 16x" above.
Repeat the cycle.
If the problem persists, call service.
Bacteriological filter clogged. Check and replace if
necessary.
19x
Clean the door seal and the chamber face side.
Repeat the cycle.
If the problem persists, call service.
2xx
Switch the sterilizer OFF and ON.
Repeat the cycle.
If the problem persists, call service.
Wait for the chamber to cool down. Reset the thermal
overload (see "Extraordinary maintenance" on page 92).