background image

39

Ensure that the parts are not contaminated on disposal. 

Follow your local and national laws, directives, standards and guidelines for disposal. 

 

> Medical device

 

> Packaging

13. Disposal

Содержание Assistina One

Страница 1: ...Instructions for Use MB 301...

Страница 2: ...ide 14 6 Start up 15 Inserting the swivel window 15 Inserting removing a filter 16 Inserting removing a cartridge 17 Connecting disconnecting the air connection hose 18 Initial filling 19 Test run 21...

Страница 3: ...31 Start up after cleaning 32 9 Maintenance 33 Replacing O rings 33 Replacing the air filter 34 10 W H Accessories and spare parts 35 11 Servicing 37 12 Technical data 38 13 Disposal 39 Explanation of...

Страница 4: ...WARNING risk of injury ATTENTION to prevent damage occurring General explanations without risk to persons or property CE mark Do not dispose of with domestic waste Symbols in the Instructions for Use...

Страница 5: ...of manufacture Do not dispose of with domestic waste Data Matrix code for product information including UDI Unique Device Identification Catalogue number Serial number CE mark Symbols on the medical...

Страница 6: ...h care Keep dry Permitted temperature range for storage Humidity limitation CE mark Keep away from heat Use by Permitted temperature range for transport 40 C 104 F Max 5 C 41 F Min Storage 50 C 122 F...

Страница 7: ...riven dental scalers Misuse may damage the medical device and hence cause risks and hazards for user and third parties When transmission instruments are prepared with the medical device they are neith...

Страница 8: ...O rings cartridges filter and air filter Assembly modifications or repairs must only be undertaken by an authorized W H service partner see page 41 Unauthorized opening of the equipment invalidates al...

Страница 9: ...disposed of by industrial recycling companies However we recommend that you keep the original packaging Remove the inserts Lift out the medical device and swivel window 2 Unpacking Remove the cardboar...

Страница 10: ...EF 30398002 REF 30398003 REF 02697000 Air connection hose X X X REF 08064970 One Care SET MB 301 W H Activefluid cartridge MC 302 200 ml W H Service Oil F1 cartridge MD 302 200 ml filter X X X REF 064...

Страница 11: ...f with soap and plenty of water Following contact with eyes Rinse opened eye for several minutes under running water If symptoms persist consult a doctor If swallowed If symptoms persist consult a doc...

Страница 12: ...r W H Activefluid MC 302 Fill level indicator W H Service Oil F1 MD 302 Viewing window for ball W H Activefluid MC 302 Viewing window for ball W H Service Oil F1 MD 302 Start button Air duct Cartridge...

Страница 13: ...13 Description Quick ISO Quick RM adaptor Quick RM adaptor Quick ISO adaptor O rings Union nut Release button O rings...

Страница 14: ...14 Connecting socket for air connection hose Abdeckung Filterfach Description Rear panel underside Cartridge compartment cover Filter compartment cover...

Страница 15: ...15 Insert the swivel window 6 Start up Inserting the swivel window...

Страница 16: ...ructions on the use of disinfectant Start up Inserting removing a filter Check the filter for damage e g vents seal Only perform the maintenance process with a functional filter Replace the filter com...

Страница 17: ...ridge compartment cover Inserting a cartridge Push the cartridge in until the limit stop is reached Removing a cartridge Pull out the cartridge Replace the cartridge compartment cover Whenever you rep...

Страница 18: ...d oil free dry and filtered compressed air according to ISO 7494 2 Replace the filter if contaminated Start up Connecting disconnecting the air connection hose Connect the air connection hose Disconne...

Страница 19: ...ut a Quick ISO Quick RM adaptor and transmission instrument Start up Initial filling Close swivel window Press the Start button The start button returns to its initial position with an audible click P...

Страница 20: ...tion of the start button A green ball appears in each of the two viewing windows After a few seconds both balls will sink again The initial filling has completed successfully If both balls are not vis...

Страница 21: ...instrument will be slightly moist when you take it out It then needs to be manually disinfected and sterilized In the event of operating malfunctions e g vibrations unusual noise overheating failure...

Страница 22: ...22 Wear protective clothing safety glasses face mask and gloves Refer to the hygiene and maintenance instructions in the instructions for use of the transmission instruments 7 Operation General...

Страница 23: ...nsmission instrument onto the adaptor with the coolant outlet facing downwards For air motors set the direction of rotation via the switching ring to clockwise or anti clockwise No maintenance is carr...

Страница 24: ...ck ISO adaptor Removing the transmission instruments with ISO connection Press the release button and pull off the transmission instrument with adaptor Tip the release button on the adaptor forwards a...

Страница 25: ...e by hand Push on the Quick RM adaptor with transmission instrument or if applicable the adaptor with transmission instrument Removing of the transmission instruments and or adaptors with fixed RM con...

Страница 26: ...ess the Start button The maintenance process runs automatically and takes approx 20 seconds A green ball appears in each of the two viewing windows The balls sink again The maintenance process has com...

Страница 27: ...y instrument Use the Quick ISO adaptors REF 06422600 and REF 02693000 Insert the adaptor with the maintenance attachment facing downwards Operation Lubrication of the chucking system Align the instrum...

Страница 28: ...ctive clothing safety glasses face mask and gloves Aerosol mist escapes The balls sink again The lubrication has completed successfully Press the Start button The lubrication runs automatically and ta...

Страница 29: ...disinfection W H recommends cleaning the medical device at least once daily or as required Wear protective clothing safety glasses face mask and gloves Cleaning agents Use only detergents which are i...

Страница 30: ...30 Disconnect the air connection hose Cleaning the Assistina One w Push swivel window upwards Remove swivel window Remove swivel window Preparation Close swivel window...

Страница 31: ...of the interior and exterior Rinse the sliding window off with running tap water 35 C 95 F and wipe down with disinfectant Manual cleaning of the interior and exterior Clean the chamber with a dry clo...

Страница 32: ...32 Start up after cleaning Position the swivel window Connect the air connection hose Cleaning the Assistina One Insert the swivel window...

Страница 33: ...w O ring Replace damaged or leaking O rings immediately Do not use sharp tools Firmly squeeze the O ring between your thumb and index finger so that it forms a loop Pull off the O ring 9 Maintenance R...

Страница 34: ...he central air supply before replacing the air filter Remove the air connection hose Push the ring a b towards the air filter and detach the air hose Replace the air filter and reconnect the air hose...

Страница 35: ...re parts or accessories approved by W H 10 W H Accessories and spare parts 06395100 Quick RM adaptor 06422600 Quick ISO adaptor 02693000 Adaptor for lubricating the chucking system For other adapters...

Страница 36: ...large 02060200 O Ring Roto Quick adaptor small 03006100 O Ring Quick RM and Quick ISO adaptor large 01208300 O Ring Quick RM and Quick ISO adaptor small 02695700 O ring Quick ISO adaptor 02675200 Air...

Страница 37: ...s In the event of operating malfunctions immediately contact an authorized W H service partner Repairs and maintenance work must only be undertaken by an authorized W H service partner Always return e...

Страница 38: ...ume 60 dB Dimensions in mm WxDxH 166 x 383 x 243 mm Weight 2 2 kg Capacity 200 ml W H Activefluid MC 302 200 ml W H Service Oil F1 MD 302 12 Technical data Ambient conditions Temperature during transp...

Страница 39: ...39 Ensure that the parts are not contaminated on disposal Follow your local and national laws directives standards and guidelines for disposal Medical device Packaging 13 Disposal...

Страница 40: ...ble for material or manufacturing defects within a warranty period of 12 months from the date of purchase We accept no responsibility for damage caused by incorrect handling or by repairs carried out...

Страница 41: ...41 Authorized W H service partners Find your nearest W H service partner at http wh com Simply go to the menu option Service for full details Or simply scan the QR code...

Страница 42: ...Manufacturer W H Dentalwerk B rmoos GmbH Ignaz Glaser Stra e 53 5111 B rmoos Austria t 43 6274 6236 0 f 43 6274 6236 55 office wh com wh com Form Nr 51046 AEN Rev 000 11 05 2022 Subject to alterations...

Отзывы: