Remove the adaptor.
Operation
Lubrication of the chucking system
Wear protective clothing, safety glasses, face mask and gloves.
Aerosol mist escapes.
> The balls sink again.
> The lubrication has completed successfully.
Press the Start button.
> The lubrication runs automatically and
takes approx. 20 seconds.
> A green ball appears in each of the two viewing windows.