Waltron User Manual 101-054-A.1 
9072 Hydrazine Analyzer 

MAINTENANCE 

6

 

MAINTENANCE 

 

6.1

 

CLEANING OF ELECTRODES 

Reference electrode: The surface of the reference electrode is made of silver chloride 
(AgCl), by a galvanic process, and must not be cleaned.  
 
Measuring electrode: The measuring electrode has to be cleaned (activated) if the 
sensitivity (k-Factor) of the measuring electrode is beneath 10 % of the determined value, 
calculated after that calibration, which followed a cleaning. As discussed above, the k-
Factor can be looked up in the menu Parameter > in the 7

th

 level. 

 
Before you remove the measuring electrode, it is necessary to empty the analyzer (of 
sample). Then remove the plug (M) from the measuring electrode and unscrew the thrust 
screw of the electrode. Take the electrode slowly out of the cell. You should be careful 
not to drop the measuring electrode and not to hit it against the cell wall. Clean the 
surface with a mild detergent, by rubbing gently the electrode between your middle 
finger, index finger and thumb. During the cleaning, do not press too hard the electrode. 
This procedure should not last longer than 2 minutes (the Pt-layer is only several µ m 
thick). Rinse well with water and do not touch the surface of the electrode with your 
fingers thereafter. 
 
Occasionally a wiping of the electrode surface with a clean tissue is sufficient. Thereby 
you reduce the later start up time until steady state condition. 
 

6.2

 

CLEANING OF CELLS 

The cleaning cycle of both the counter electrodes of the measuring and calibration cells is 
the same as those of the measuring electrode and calibration electrode. To clean the 
counter electrodes, you close the inlet valve and remove the plug connectors (A), (K) for 
the calibration cell, as well as (M), (C) and (R) for the measuring cell. Disconnect the 
pipe between the calibration cell and the flowmeter. Disassemble the calibration cell from 
the square mounting by detaching the fitting (screw width SW 12 mm). Remove the 
calibration electrode (see also the instruction for cleaning electrodes above). Clean the 
interior with a mild detergent and a thin test tube brush (see also “Spare parts“). Examine, 

if the nozzle in the top part (diameter 

 0.8 mm) is free. After the cleaning you have to 

rinse the cell well with water. The dismantling of the measuring cell from the square 
mounting should be performed in the same way as the calibration cell. Remove the outlet 
pipe and unscrew the reference cell from the measuring cell (see also 

Section 4.2

). Take 

the electrode slowly out of the cell (see also 

Section 6.1

).  The cleaning procedure 

thereafter is the same, as described for the calibration cell.  
 
When reassembling, make sure that all parts are dry. 
 
 

Содержание 9072

Страница 1: ...101 054 A 1 EXPERTS IN WATER CHEMISTRY SINCE 1903 9072 Hydrazine Analyzer User Manual...

Страница 2: ...n net Please be ready to provide the following information Date analyzer was purchased Analyzer model and serial number Recent maintenance history Calibration slope values and detailed description of...

Страница 3: ...ontainers packages and chemicals Only qualified personnel should be involved in the installation operation and servicing of the analyzer Follow safety precautions when operating analyzer in conditions...

Страница 4: ...ce Department Customer name address and department Name and telephone number of the individual responsible for returning items for repair Brief problem description Ship to Waltron service center Waltr...

Страница 5: ...ping needs Check all items received against those on the packing list Chemicals are usually shipped in a separate package and will be itemized accordingly Verify that the number of packages received a...

Страница 6: ...ED PARTS 10 2 2 ACCESSORIES 14 2 3 TERMINAL SCHEMATIC 14 2 4 TECHNICAL DATA 15 3 OPERATING PROCEDURES 16 3 1 PRINCIPLES OF MEASREMENT 16 4 START UP 16 4 1 MENU STRUCTURE 16 4 2 PREPARING TO MEASURE 18...

Страница 7: ...reduce the dissolved oxygen concentrations to protect iron containing metallurgy Hydrazine concentration needs to be continuously monitored to provide feedback to the chemical injection systems and t...

Страница 8: ...is the analytical component which is composed of the inlet valve filter device and flow meter which serve the sample and probe preparation The measuring cell is joined with the temperature sensor and...

Страница 9: ...8 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INTRODUCTION 1 3 THE MEASURING SYSTEM...

Страница 10: ...9 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INTRODUCTION 1 4 BLOCK DIAGRAM OF SENSOR SYSTEM...

Страница 11: ...anual 101 054 A 1 9072 Hydrazine Analyzer INSTALLATION 2 INSTALLATION 2 1 DELIVERED PARTS Arrangement of the wall mounted instrument M measuring electrode C counter electrode R reference electrode T t...

Страница 12: ...11 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INSTALLATION TRANSMITTER MOUNTING...

Страница 13: ...12 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INSTALLATION WET SECTION MOUNTING...

Страница 14: ...8 9 Line 100 230 V 50 60 Hz Line 100 230 V 50 60 Hz Ground earth Limit alarm off Limit alarm on Limit alarm common Alarm on Alarm off Alarm common Auto changing range 2 Auto changing range 1 Auto chan...

Страница 15: ...ime buffer 21 000 data points time set 1 s to 5 min Limit power relay Alarming power relay Flow and Calibration Measuring electrode gold Counter electrode stainless steel 1 4571 314 Reference Electrod...

Страница 16: ...ith that of the reference electrode Under these conditions the rate of the electrochemical reaction is limited by the rate of diffusion of hydrazine molecules to the electrode As long as we remain wit...

Страница 17: ...12 00 SET Menu 03 11 RangeModus 1 2 AutoCal 2 3 AutoCal Period 1 3 RangeModus ManRange SET Period daily RangeModus 2 2 AutoCal Period 2 3 AutoRange SET weekly Menu 04 11 I Output AutoCal Period 3 3 I...

Страница 18: ...Interval 2 5 30 sec SET Logging Interval 3 5 1 min SET Logging Interval 4 5 2 min SET Logging Interval 5 5 5 min SET Logging 3 4 Logging Control Control Press to stop logging Logging Control press to...

Страница 19: ...to the inlet valve The outlet line is placed at the measuring cell exit with a hose having a 5 mm inside diameter 4 3 CHOICE OF OPERATING PARAMETERS Before starting measurement set your operating par...

Страница 20: ...ut current 0 4 20 mA 5th Menu level Limit allows limit settings There exists a switch hysteresis of 10 of the chosen limit 6th Menu level FlowCal enables flow meter calibration see Section 7 7th Menu...

Страница 21: ...verify that all air bubbles in the tube system rinse out To help the bubbles to rinse out open and close quickly the inlet valve several times Attention Make sure that the plugs stay dry 5 2 ANALYZER...

Страница 22: ...ml solution A to 1000 ml deionized water 150 g l N2H4 Calibration solution Add 16 7 ml of solution B to 5 l hydrazine free water 500 g l N2H4 You start the calibration procedure by pressing key and i...

Страница 23: ...de with your fingers thereafter Occasionally a wiping of the electrode surface with a clean tissue is sufficient Thereby you reduce the later start up time until steady state condition 6 2 CLEANING OF...

Страница 24: ...ecting the flow 4 The amount in ml of sample collected in the cylinder during the 60 seconds should be entered digit by digit with the keys and If the value is correct press key SET Press key ESC to g...

Страница 25: ...0 009 1 Measuring electrode Pt H2 L1000 010 1 Salting cell TA L1000 011 1 Salting cell TI L1000 012 1 Sieve L1000 013 1 Set of O rings TI L1000 014 1 Flow meter TA L1000 015 1 Flow meter TI L1000 016...

Отзывы: