background image

    

 

 

 

                                   

 

 

 

 

15 

Waltron User Manual 101-054-A.1 

9072 Hydrazine Analyzer 

OPERATING PROCEDURES 

3

 

OPERATING PROCEDURES 

 

3.1

 

PRINCIPLE OF MEASUREMENT 

The 9072 Hydrazine Analyzer electrochemical hydrazine sensor functions in a 
potentiostatic mode with an open three-electrode system. The measuring electrode is 
made of solid titanium, coated with gold. The counter electrode is made of stainless steel 
and the reference electrode is made of silver/silver-chloride. The use of open measuring 
electrode means, that there is a direct contact between the sample and the measuring 
electrode. A membrane is not required. For the measurement of dissolved hydrazine 
(N2H4) the fluid medium flows into the coaxial measurement cell between the platinum 
measuring electrode and counter electrode. If the measuring electrode has reached the 
correct potential, hydrazine is oxidized in the interphase boundary layer. The 
electrochemical reaction may represented qualitatively as:  
 

4OH- 

 N

+ H

2

O + 4e-  

  

The characteristic potential for this reaction the potential at which hydrazine oxidation occur 
in preference to other competing reactions is measured by holding the potential of the 
measuring electrode steady with a potentiostatic and comparing its value with that of the 
reference electrode. Under these conditions the rate of the electrochemical reaction is limited 
by the rate of diffusion of hydrazine molecules to the electrode. As long as we remain within 
this ‘diffusion-limiting-current’ area, the electrochemical cell delivers an electric current, 
whose size depends linearly on the concentration of dissolved hydrazine. This current, which 
is measured by the instrument’s electronic processing unit, is known as the diffusion-limiting 
current,

  

 
I

diff

: cN2 H4 = k • I

diff

 

 

 
This diffusion-limiting current depends mainly on the mass transport of the 
electrochemically active components, and hence on the hydrodynamics in the measuring 
cell. It therefore changes significantly with the temperature and flowrate through the 
measuring cell. In practice the flowrate and temperature vary over wide ranges, but the 
instrument corrects these changes by using empirical correlations. 
 
A second important influence on the measurement is the condition of the electrodes. In 
particular the adsorptive and oxidative state of the electrode surface layers. This is 
independent of the flow conditions; it depends instead on the voltage between the 
electrode and the electrolyte, and on the material of which the electrode is made 
Appearing flow- and temperature effects of the measuring signal are recorded with the 
aid of the installed flow meter and temperature sensor and are compensated, if necessary.  
 
 
 
 
 

Содержание 9072

Страница 1: ...101 054 A 1 EXPERTS IN WATER CHEMISTRY SINCE 1903 9072 Hydrazine Analyzer User Manual...

Страница 2: ...n net Please be ready to provide the following information Date analyzer was purchased Analyzer model and serial number Recent maintenance history Calibration slope values and detailed description of...

Страница 3: ...ontainers packages and chemicals Only qualified personnel should be involved in the installation operation and servicing of the analyzer Follow safety precautions when operating analyzer in conditions...

Страница 4: ...ce Department Customer name address and department Name and telephone number of the individual responsible for returning items for repair Brief problem description Ship to Waltron service center Waltr...

Страница 5: ...ping needs Check all items received against those on the packing list Chemicals are usually shipped in a separate package and will be itemized accordingly Verify that the number of packages received a...

Страница 6: ...ED PARTS 10 2 2 ACCESSORIES 14 2 3 TERMINAL SCHEMATIC 14 2 4 TECHNICAL DATA 15 3 OPERATING PROCEDURES 16 3 1 PRINCIPLES OF MEASREMENT 16 4 START UP 16 4 1 MENU STRUCTURE 16 4 2 PREPARING TO MEASURE 18...

Страница 7: ...reduce the dissolved oxygen concentrations to protect iron containing metallurgy Hydrazine concentration needs to be continuously monitored to provide feedback to the chemical injection systems and t...

Страница 8: ...is the analytical component which is composed of the inlet valve filter device and flow meter which serve the sample and probe preparation The measuring cell is joined with the temperature sensor and...

Страница 9: ...8 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INTRODUCTION 1 3 THE MEASURING SYSTEM...

Страница 10: ...9 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INTRODUCTION 1 4 BLOCK DIAGRAM OF SENSOR SYSTEM...

Страница 11: ...anual 101 054 A 1 9072 Hydrazine Analyzer INSTALLATION 2 INSTALLATION 2 1 DELIVERED PARTS Arrangement of the wall mounted instrument M measuring electrode C counter electrode R reference electrode T t...

Страница 12: ...11 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INSTALLATION TRANSMITTER MOUNTING...

Страница 13: ...12 Waltron User Manual 101 054 A 1 9072 Hydrazine Analyzer INSTALLATION WET SECTION MOUNTING...

Страница 14: ...8 9 Line 100 230 V 50 60 Hz Line 100 230 V 50 60 Hz Ground earth Limit alarm off Limit alarm on Limit alarm common Alarm on Alarm off Alarm common Auto changing range 2 Auto changing range 1 Auto chan...

Страница 15: ...ime buffer 21 000 data points time set 1 s to 5 min Limit power relay Alarming power relay Flow and Calibration Measuring electrode gold Counter electrode stainless steel 1 4571 314 Reference Electrod...

Страница 16: ...ith that of the reference electrode Under these conditions the rate of the electrochemical reaction is limited by the rate of diffusion of hydrazine molecules to the electrode As long as we remain wit...

Страница 17: ...12 00 SET Menu 03 11 RangeModus 1 2 AutoCal 2 3 AutoCal Period 1 3 RangeModus ManRange SET Period daily RangeModus 2 2 AutoCal Period 2 3 AutoRange SET weekly Menu 04 11 I Output AutoCal Period 3 3 I...

Страница 18: ...Interval 2 5 30 sec SET Logging Interval 3 5 1 min SET Logging Interval 4 5 2 min SET Logging Interval 5 5 5 min SET Logging 3 4 Logging Control Control Press to stop logging Logging Control press to...

Страница 19: ...to the inlet valve The outlet line is placed at the measuring cell exit with a hose having a 5 mm inside diameter 4 3 CHOICE OF OPERATING PARAMETERS Before starting measurement set your operating par...

Страница 20: ...ut current 0 4 20 mA 5th Menu level Limit allows limit settings There exists a switch hysteresis of 10 of the chosen limit 6th Menu level FlowCal enables flow meter calibration see Section 7 7th Menu...

Страница 21: ...verify that all air bubbles in the tube system rinse out To help the bubbles to rinse out open and close quickly the inlet valve several times Attention Make sure that the plugs stay dry 5 2 ANALYZER...

Страница 22: ...ml solution A to 1000 ml deionized water 150 g l N2H4 Calibration solution Add 16 7 ml of solution B to 5 l hydrazine free water 500 g l N2H4 You start the calibration procedure by pressing key and i...

Страница 23: ...de with your fingers thereafter Occasionally a wiping of the electrode surface with a clean tissue is sufficient Thereby you reduce the later start up time until steady state condition 6 2 CLEANING OF...

Страница 24: ...ecting the flow 4 The amount in ml of sample collected in the cylinder during the 60 seconds should be entered digit by digit with the keys and If the value is correct press key SET Press key ESC to g...

Страница 25: ...0 009 1 Measuring electrode Pt H2 L1000 010 1 Salting cell TA L1000 011 1 Salting cell TI L1000 012 1 Sieve L1000 013 1 Set of O rings TI L1000 014 1 Flow meter TA L1000 015 1 Flow meter TI L1000 016...

Отзывы: