safe place. Cut away the plastic film around the unit and remove the four polystyrene 
support corners. Do not pick up the unit by the fascia panel edges but support it below the 
white cabinet. 

(d) Using a RESIST-TORX T30 security key undo the two screws on the unit top fascia panel 

and lift the panel off. Place it in a safe place resting on its rear face. 

(e) Release the three screws securing the inner top panel metal cover & remove panel. If 

more than one unit is being installed, ensure the panels are kept with each unit. 

(f)  Using the other key type provided, unlock the lower drawer locks (2). Pull open the drawer 

& remove the plastic soap container & hook over the soap tube and priming switch into 
the wash basin on the unit. 

(g) Remove the drawer from the unit by pressing the black plastic levers that are located on 

each slide. Note that the left lever should be pressed down while the right lever is lifted 
up. Pull the drawer out from the unit until it is released from the slides. This operation may 
require two people or alternately stand the unit up on its base with a pad below it to raise 
the drawer off floor level. 

(h) Move the unit to the aperture and pass the electrical supply cable through the entry hole 

on the back of the cabinet (located towards left-hand side). 

(i)  Lift the unit up into the aperture and secure to the mounting plates with the bolts & 

washers provided. Tighten (hand tight only) the bolts half way into the bushes only at this 
stage. 

(j)  Adjust the brass nuts on the bushes by unscrewing them so that they meet the sides of 

the cabinet. The nuts must be adjusted evenly each side so that the cabinet is positioned 
centrally in the aperture. Extra care must be taken with the two lower points to avoid 
distorting the cabinet sides in too far which will then prevent the satisfactory operation of 
the drawer in its slides. The cabinet sides should be set apart a distance of 541mm (21 
5/16”) to their outer edges. Once all four points are set, press the unit fully into the 
aperture and tighten the four bolts. 

 

4. ELECTRICAL CONNECTION. 

WARNING: This appliance must be earthed. 

(a) The CME is designed for a fixed wire type installation. 
(b) The unit should be installed by a qualified electrician in accordance with the current 

Nationally Approved Regulations (IEE). 

(c) Terminate the supply cable into the terminal block positions “L E N” located on the left 

side within the cabinet and secure it within the cable clamp. The supply cable must not 
come into contact with any other components or wiring within the unit. 

(d) Earth bonding should be carried out to comply with the Nationally Approved Regulations 

(IEE). Connections must be made to the following points: Water supply pipe, Sink waste 
outlet & Equipotential earth connection for bonding together multiple units installed, 
(connection positioned on back of cabinet behind drawer). 

 

5. WATER SUPPLY & WASTE CONNECTION. 

IMPORTANT!

 For installations liable to freezing conditions, frost protection & pipe insulation 

must be fitted. 
(a) Prior to final connection of the water supply to the unit, the pipes must be flushed through 

so as to remove any debris & air locks. 

Warning: Failure to carry this out may result in damage and failure of the unit. 

(b) Complete the water connection to the plastic tap/flow regulator on the unit, taking care 

that any sealing compounds used are not excessively applied, which may cause a 
blockage. 

(c) For the waste connection an appropriate trap (not supplied, recommend McAlpine type 

SB10 or SP10) should be fitted to the units sink waste (1 ¼” BSP). Ensure the trap is 
positioned only on the centre line of the unit so that it does not interfere with the drawer 
unit. Final connection can then be made to the main waste pipe. 
 

Содержание CME 41

Страница 1: ...ts Contact our Customer Services Department Tel 01722 744594 Fax 01722 742096 Sales or 742933 Service Email sales wallgate com service wallgate com Crow Lane Wilton Salisbury Wiltshire SP2 OHB www wallgate com Hand Wash Drier Models CME_41 45 Product Manual CME_45 ...

Страница 2: ...st be installed c Complete all surface finishes i e screed plaster or tiling to the wall before fixing the mounting brackets as these must be fitted flush with the finished surface d An electrical supply should be provided with a switched fused spur 13Amp double pole connection point that is local to the unit the contact separation of the fused spur switch should be at least 3mm A suitable cable w...

Страница 3: ...et sides in too far which will then prevent the satisfactory operation of the drawer in its slides The cabinet sides should be set apart a distance of 541mm 21 5 16 to their outer edges Once all four points are set press the unit fully into the aperture and tighten the four bolts 4 ELECTRICAL CONNECTION WARNING This appliance must be earthed a The CME is designed for a fixed wire type installation...

Страница 4: ...bove the basin The isolator switch should illuminate red f To check that the unit operates and to complete the final set up procedures now refer to the Operating Maintenance Instructions sections 3 2 4 in this order g On completion of the above replace the top fascia panel and lock the lower drawer The following items should be passed to the customer 1 off lower drawer key this Product Manual 7 TE...

Страница 5: ......

Страница 6: ...over due to the risk of electric shock For the CME 41 push button operated model follow the instructions below For soap dispensing press the hand or foot button marked SOAP Dosage depends entirely on the number of times the button is pressed The electric type pump runs for a pre set time 0 3 seconds to deliver approximately 5ml of soap This can only be adjusted by consulting the manufacturer The s...

Страница 7: ...with one of two electronic controllers TCU041 or TCU100 which is mounted towards the top inside face of the handwash housing The TCU041 controller is used on units with just soap water and air functions push button operated The time settings may be adjusted by following the adjacent instructions The TCU100 is fitted to the CME_45 model time cycles may be adjusted using the onboard Control Interfac...

Страница 8: ...will only operate the water heater when the flow rate is set above 1 2ltr min 5 MAINTAINANCE Wallgate operates a comprehensive service package details of which can be obtained from our After Sales Service department A full spare parts facility is also available WARNING The water heater is an open outlet type and under no circumstances should the water outlet nozzle be allowed to become blocked or ...

Отзывы: