Once the parameters have been set, the scaling is automatically enabled. To disable it, enter the
Configuration/Scaling ON
menu, select
OFF
and confirm by pressing the
VAL
key.
D.2
Differential measurements
The relative measurement function available on the device makes it possible to cancel a constant or spurious
value via programming. Selecting
Configuration/NULLING/ON
will null out the value defined in the Tare
value box. Selecting
Tare
will null the current value displayed and will update the
NULLING/Define...
Tare
value box with this new value. Selecting
Configuration/NULLING/OFF
will return the unit to its original
display value.
When the relative measurement function is enabled, the symbol
is displayed on the measurement
screen.
Scaling
pictogram
Pictogram of the
TARE mode
TM612
- 18 -
The
NULLING/Define
menu makes it possible to program the value of the Tare (positive or negative). This
value is obtained from the measurements:
Value Displayed = Value measured – Value of the Tare
D.3
Calibrated sensors
The calibrated sensors function of the device makes it possible to use sensors, the calibration (correction)
coefficients of which are taken into consideration by the device during measurement.
•
Using the F1 key, enter the Configuration menu.
•
Select the
Calibrated sensors
function, followed by one of the 5 available sensors.
•
Confirm by pressing
VAL
.