TM612
- 21 -
- 21 -
The drop-down list displays the following functions:
SAVE MEASUREMENT:
Enable the triggering of an acquisition on a case-by-case basis.
If an acquisition has already been opened, then the following screen is displayed:
Press the F2 key (YES) to confirm.
•
You are then requested to enter the name of a file. Using the navigation keys (
↑
↑
↑
↑
and
↓
↓
↓
↓
), scroll
down the letters.
•
Using the navigation keys (
and
), move the cursor by one position.
•
Using the F2 key(
), you can delete the characters entered
•
Once you have entered the file name, confirm by pressing the
VAL
key.
TM612
- 22 -
RUN:
Launches the storage of data following the parameters set in the “parameters” function. The pictogram is
displayed on the measurement screen
STOP:
Stops the storage in progress.
PARAMETERS:
Allows you to define:
•
The size of the acquisition (max 10,000 values),
•
The sampling period from 0.5 S to 30 Min,
•
The type of trigger (None, low level, high level).
If you have selected a low level or high level trigger, you must define the trigger level and the number of data
to record after this trigger (Post-trigger).
Display burst:
You can display the burst in the form of a table of values or a trend curve.
Post trigger = 900 measurements
Size of block (1,000 measurements)
Trigger
on
programmed
value= 10 °C