If you prefer to calibrate it yourself and have the appropriate equipment, the temperature
controller has Low and High temperature offsets available for calibration of the internal
temperature sensor, which drives the PV display. Section 7.3 of this manual shows the controllers
registers for A.L.P, the low temperature calibration point, A.L.o, the low temperature point offset
(or correction); A.H.P, the high temperature calibration point and A.H.o, the high temperature
point offset (or correction).
For a complete example, you can download the user manual from the Ascon Technologic website
at the following link. Manuals are at the bottom of the page.
https://www.ascontecnologic.com/en/industrial-automation/advanced-programmer-
controllers/k-series/k39
10.
Troubleshooting
Failure Code /
Symptom
Possible Cause
Action to take
Displays “- - - “
The reference sensor is open or
defective
Return the instrument to the
manufacturer for repair
Displays “uuu”
Under range indication -
Reference sensor is below
range
Increase the temperature or return
the instrument to the manufacturer
for repair
Displays “ooo”
Over range indication -
Reference sensor is open or
over range
Decrease the temperature or return
the instrument to the manufacturer
for repair
Displays “EREP”
Possible failure in the EEPROM
memory of the controller
Press the “P” key. Power off unit
and restart
Fan does not work
Fan is blocked or defective
Clear any obstructions. Return the
instrument to the manufacturer for
repair
PV does not reach SV
The solid state relay is defective
or the heating/cooling elements
are defective
Return the instrument to the
manufacturer for repair
No Display
The controller is defective
Return the instrument to the
manufacturer for repair
No Functionality at all
AC power not connected
Fuse blown
Check power connections and
check the fuse
ATTENTION!
If the defect cannot be corrected by the actions detailed above the instrument must
be put out of service immediately and prevented from erroneous powering up. In
this case you must consult the manufacturer. If you wish to return the instrument
please see section 11.2 Returns.