Cooling the portable Dry Well Calibrator
WARNING! – RISK OF BURNS
Before transporting or touching the portable Dry Well Calibrator it is necessary to
ensure that it is sufficiently cooled, otherwise there is a risk of burns to the operator
and possible damage to the portable Dry Well Calibrator. To bring the calibrator
from a high temperature to a low temperature as quickly as possible, the set point
temperature must be set to a temperature lower than or near ambient temperature.
ATTENTION!
If AC Power is removed from the unit, neither the active cooling nor the fan will
operate, which will slow the cooling significantly. However, there is sufficient
insulation to ensure no damage to the Dry Well Calibrator or its surroundings will
occur.
Press and release “P” and the current
Set Point will show.
Using the up and down arrows, scroll
to the desired new set point value. In
this example, 40.0° is used. Press P to
confirm and save.
Press U for 5 seconds or wait for 10
seconds with no keys pressed and the
meter will return to normal operation.
Press
“
P”
Press up or down arrows
Press “P” to save
Press “U” for 5 seconds to
return to operating mode