www.vxitech.com
CT-100B Installation
15
P
OWER
S
PECIFICATIONS
U
SEABLE
P
OWER
450 W at 40°C
DC Supply Voltage
Voltage
Peak Current
(I
MP
)
Dynamic Current
(I
MD
)
+5
V 35
A 5
A
-5.2
V 25
A 4
A
-2
V 7
A 2
A
+12
V 4
A 2
A
-12
V 4
A 1.5
A
+24
V 4
A 3
A
-24
V 4
A 3
A
Output Voltage
Voltage
Allowed Variation
Ripple/Noise
DC Load
Induced Ripple
Noise
+5 V
+0.25 V / -0.125 V
50 mV
50 mV
-5.2 V
-0.26 V / +0.125 V
50 mV
50 mV
-2 V
-0.10 V / +0.72 V
50 mV
50 mV
+12 V
+0.60 V / +0.36 V
50 mV
50 mV
-12 V
-0.60 V / +0.36 V
50 mV
50 mV
+24 V
+1.20 V / -0.72 V
150 mV
150 mV
-24 V
-1.20 V / +0.74 V
150 mV
150 mV
P
OWER
I
NPUT
Input
Voltage
AC
Frequency
Fuse
115 V Operation
220 V Operation
115 VAC, nominal 90 – 132 VAC
230 VAC, Nominal 180 – 264 VAC
47 Hz – 63 Hz
10 A slow blow
5 A slow blow
+5VSTBY
6 A max, user supplied
Standard banana jack on rear panel, voltage and ground
P
OWER
S
UPPLIES
(TBD)
UL, CSA, TUV approved
Shout circuit, over-voltage, reverse voltage and thermal shutdown protection
Auto Selected
}