Objective specifications: R indicates retractable lens 

DIN Standard 

Objectives 

Magnificat

ion 

Numerical 

Aperture 

Working 

Distance 

Field of 

View with 

10X E.P. 

Plan achromatic 

5X 

0.12 

18.3mm 

3.6mm 

Plan achromatic 

10X 

0.25 

8.9mm 

1.8mm 

LWD planachro. 

20X R 

0.40 

4.8mm 

0.9mm 

LWD planachro. 

40X R 

0.60 

3.73mm 

0.45mm 

LWD planachro 

50X R 

0.70 

2.02mm 

0.36mm 

 

 

 

Setup Instructions 

1.

 

Remove all parts from their packing materials and retain the packaging in 

the event you need to transport the product. 

2.

 

Loosen the thumbscrew and install the Optical Body with Vertical 

Illumination Tube to the frame by removing its dust cap and inserting the 

dovetailed flange on the underside of the optical body into the frame. 

Secure optical body with thumbscrew. DO NOT RELEASE THE 

OPTICAL BODY FROM YOUR GRASP UNTIL IT IS FIRMLY   

SECURED TO THE FRAME. 

3. 

Install the objective into the nosepiece. 

4.

 

Remove the dust caps from the eyetubes in the Optical Body and insert   

both eyepieces into the Optical Body. 

5.

 

Install the lamp house. 

6.

 

Connect the power cord to the instrument and to a suitable power supply. 

Содержание VistaVision 82026-636

Страница 1: ...Metallurgical Microscope Model 82026 636 V VW WR RT TM M V Vi is st ta aV Vi is si io on nT TM M U Us se er r G Gu ui id de e...

Страница 2: ...AIN SOCKET BEFORE REMOVING THE BOTTOM PLATE OPENING THE LAMP REPLACEMENT DOOR OR REMOVING THE LAMP HOUSE WARNING THE POWER CORD PROVIDED WITH THE EQUIPMENT HAS A GROUNDED PLUG ALWAYS USE THE POWER COR...

Страница 3: ...of Contents 3 Getting Started 3 Components 4 Technical Specifications 5 Setup Instructions 6 Basic Operation 7 Changing the Incident upper Light Bulb 13 Changing the Transmitted lower Light Bulb 13 Pr...

Страница 4: ...re Halogen Bulbs Spare Fuse Power Cord Dustcover Manual Getting Started This reference manual is written on the assumption that it will be used by advanced students and experienced microscopists It co...

Страница 5: ...13 Coarse Focusing Knob 14 Fine Focusing Knob 15 Plug for Vertical Illumination 16 Transmitted Incident Illumination Toggle Switch 17 Lengthwise Knob 18 Cross Knob 19 Collector 20 Field Diaphragm 21...

Страница 6: ...otion 75mm X 50mm Focus Mechanism All stainless steel and brass gear train system Adjustable Tension Control to prohibit drift and Adjustable Up Stop to protect slides specimen and objective lenses Co...

Страница 7: ...ng in the event you need to transport the product 2 Loosen the thumbscrew and install the Optical Body with Vertical Illumination Tube to the frame by removing its dust cap and inserting the dovetaile...

Страница 8: ...nsmitted incident switch is set on the transmitted setting 2 Focusing Controls a Focusing adjustment is accomplished by using the large coarse adjustment knobs located comfortably on each side of the...

Страница 9: ...ts use also simplifies focusing After focusing on the specimen with the coarse adjustment rotation of the lever toward the front of the microscope will set the upper limit on the coarse adjustment mov...

Страница 10: ...follows see Figure 6 Abbe condenser its Mounting Ring with lock screw and two large centering screws at either side for adjusting condenser iris assembly in the optical center of the light path the a...

Страница 11: ...A This white circle is the light that is passing through the iris and it should be centered in the black field as in Figure 8b If it is not centered move it to the center using the centering screws iv...

Страница 12: ...Reduce the aperture iris diaphragm using the adjustment lever and rotate the lamp focusing control on the lamp housing until a sharp image of the lamp filament is attained 4 Refer to Figure 6 for thi...

Страница 13: ...ity filters as well as a clear opening 10 The Polarizing Analyzer is used in conjunction with the Polarized filter cube selected by pushing in the blue filter rod Figure 10 The Analyzer may also be pl...

Страница 14: ...from reducing the bulb s intensity and life 4 Reinsert the lamp socket into the lamp house and retighten the setscrew Make certain that the new lamp is placed squarely in the socket to insure that is...

Страница 15: ...on Be sure the instrument is dry before using 2 Cleaning optical parts Microscope eyepieces and objectives are coated They should not be wiped while dry as dirt or dust may scratch the coating It is b...

Страница 16: ...15 VWR TECHNICAL HOT LINE 888 VWR LINE 888 897 5463...

Отзывы: