background image

 

10 

a)

 

Centering

–The condenser must be 

centered in the light path to ensure 

proper light control. One simple 

method for centering follows: 

i) Rotate the objective nosepiece until        

the 5X objective is in the light path. 

ii)  While looking into the microscope,       

close the field diaphragm to the 

smallest setting using the knurled ring 

at the top of the light exit window 

assembly. 

iii) Closing the iris in this manner will 

reduce the field so that a small white 

circle is visible within a black field, see 

Figure 8A. This white circle is the light 

that is passing through the iris and it should be centered in the black field 

as in Figure 8b. If it is not centered, move it to the center using the 

centering screws. 

iv) By opening the iris until the white circle almost fills the entire field, 

see Figure 8C. 

b)

 

Vertical Focusing–

The condenser can be raised and lowered with the 

sub-stage adjustment knob in order to focus the light for optical 

illumination. 

c)

 

Aperture Adjustment–

The contrast can be adjusted with the iris 

diaphragm adjustment lever on the condenser assembly. Aperture 

adjustments are made to induce contrast into a specimen, not to adjust 

light intensity. 

 

 

 

 

Figure 8 

Substage 

Adjustment Knob 

Figure 7 

Содержание VistaVision 82026-636

Страница 1: ...Metallurgical Microscope Model 82026 636 V VW WR RT TM M V Vi is st ta aV Vi is si io on nT TM M U Us se er r G Gu ui id de e...

Страница 2: ...AIN SOCKET BEFORE REMOVING THE BOTTOM PLATE OPENING THE LAMP REPLACEMENT DOOR OR REMOVING THE LAMP HOUSE WARNING THE POWER CORD PROVIDED WITH THE EQUIPMENT HAS A GROUNDED PLUG ALWAYS USE THE POWER COR...

Страница 3: ...of Contents 3 Getting Started 3 Components 4 Technical Specifications 5 Setup Instructions 6 Basic Operation 7 Changing the Incident upper Light Bulb 13 Changing the Transmitted lower Light Bulb 13 Pr...

Страница 4: ...re Halogen Bulbs Spare Fuse Power Cord Dustcover Manual Getting Started This reference manual is written on the assumption that it will be used by advanced students and experienced microscopists It co...

Страница 5: ...13 Coarse Focusing Knob 14 Fine Focusing Knob 15 Plug for Vertical Illumination 16 Transmitted Incident Illumination Toggle Switch 17 Lengthwise Knob 18 Cross Knob 19 Collector 20 Field Diaphragm 21...

Страница 6: ...otion 75mm X 50mm Focus Mechanism All stainless steel and brass gear train system Adjustable Tension Control to prohibit drift and Adjustable Up Stop to protect slides specimen and objective lenses Co...

Страница 7: ...ng in the event you need to transport the product 2 Loosen the thumbscrew and install the Optical Body with Vertical Illumination Tube to the frame by removing its dust cap and inserting the dovetaile...

Страница 8: ...nsmitted incident switch is set on the transmitted setting 2 Focusing Controls a Focusing adjustment is accomplished by using the large coarse adjustment knobs located comfortably on each side of the...

Страница 9: ...ts use also simplifies focusing After focusing on the specimen with the coarse adjustment rotation of the lever toward the front of the microscope will set the upper limit on the coarse adjustment mov...

Страница 10: ...follows see Figure 6 Abbe condenser its Mounting Ring with lock screw and two large centering screws at either side for adjusting condenser iris assembly in the optical center of the light path the a...

Страница 11: ...A This white circle is the light that is passing through the iris and it should be centered in the black field as in Figure 8b If it is not centered move it to the center using the centering screws iv...

Страница 12: ...Reduce the aperture iris diaphragm using the adjustment lever and rotate the lamp focusing control on the lamp housing until a sharp image of the lamp filament is attained 4 Refer to Figure 6 for thi...

Страница 13: ...ity filters as well as a clear opening 10 The Polarizing Analyzer is used in conjunction with the Polarized filter cube selected by pushing in the blue filter rod Figure 10 The Analyzer may also be pl...

Страница 14: ...from reducing the bulb s intensity and life 4 Reinsert the lamp socket into the lamp house and retighten the setscrew Make certain that the new lamp is placed squarely in the socket to insure that is...

Страница 15: ...on Be sure the instrument is dry before using 2 Cleaning optical parts Microscope eyepieces and objectives are coated They should not be wiped while dry as dirt or dust may scratch the coating It is b...

Страница 16: ...15 VWR TECHNICAL HOT LINE 888 VWR LINE 888 897 5463...

Отзывы: