VWR VistaVision 82026-636 Скачать руководство пользователя страница 13

 

12 

5)

 

Reopen the aperture iris diaphragm. 

6)

 

Reinstall the objective again.

 

7)

 

The Iris diaphragm nearest the illuminator house in the vertical 

illuminator tube may be closed slightly to enhance contrast. The iris 

diaphragm nearest the illuminator house should be centered for 

Kohler illumination.

 

8)

 

There is a field diaphragm near the trinocular body for Kohler 

Illumination.

 

9)

 

The Filter wheel contains a green, yellow, blue, and neutral density 

filters, as well as a clear opening.

 

10)

 

The Polarizing Analyzer is used in 

conjunction with the Polarized 

filter cube, selected by pushing in 

the blue filter rod (Figure 10). The 

Analyzer may also be placed over 

the lower light source on the base 

(Figure 7) and used with the 

Polarized filter cube for 

transmitted light work.

 

11)

 

The white push rod to the upper 

left of the trinocular head will 

send 100%  of the image light to 

the binocular eyepieces or to the 

top port.

 

 

 

 

 

 

 

 

 

Figure 10 

1. Camera Port    2. Push Rod to Direct 

Image to Binocular Eyepieces or 

Camera Port (White) 3.Silver Rod for 

Normal Upper Light 4.  Blue Rod for 

Analyzer Filter 

Содержание VistaVision 82026-636

Страница 1: ...Metallurgical Microscope Model 82026 636 V VW WR RT TM M V Vi is st ta aV Vi is si io on nT TM M U Us se er r G Gu ui id de e...

Страница 2: ...AIN SOCKET BEFORE REMOVING THE BOTTOM PLATE OPENING THE LAMP REPLACEMENT DOOR OR REMOVING THE LAMP HOUSE WARNING THE POWER CORD PROVIDED WITH THE EQUIPMENT HAS A GROUNDED PLUG ALWAYS USE THE POWER COR...

Страница 3: ...of Contents 3 Getting Started 3 Components 4 Technical Specifications 5 Setup Instructions 6 Basic Operation 7 Changing the Incident upper Light Bulb 13 Changing the Transmitted lower Light Bulb 13 Pr...

Страница 4: ...re Halogen Bulbs Spare Fuse Power Cord Dustcover Manual Getting Started This reference manual is written on the assumption that it will be used by advanced students and experienced microscopists It co...

Страница 5: ...13 Coarse Focusing Knob 14 Fine Focusing Knob 15 Plug for Vertical Illumination 16 Transmitted Incident Illumination Toggle Switch 17 Lengthwise Knob 18 Cross Knob 19 Collector 20 Field Diaphragm 21...

Страница 6: ...otion 75mm X 50mm Focus Mechanism All stainless steel and brass gear train system Adjustable Tension Control to prohibit drift and Adjustable Up Stop to protect slides specimen and objective lenses Co...

Страница 7: ...ng in the event you need to transport the product 2 Loosen the thumbscrew and install the Optical Body with Vertical Illumination Tube to the frame by removing its dust cap and inserting the dovetaile...

Страница 8: ...nsmitted incident switch is set on the transmitted setting 2 Focusing Controls a Focusing adjustment is accomplished by using the large coarse adjustment knobs located comfortably on each side of the...

Страница 9: ...ts use also simplifies focusing After focusing on the specimen with the coarse adjustment rotation of the lever toward the front of the microscope will set the upper limit on the coarse adjustment mov...

Страница 10: ...follows see Figure 6 Abbe condenser its Mounting Ring with lock screw and two large centering screws at either side for adjusting condenser iris assembly in the optical center of the light path the a...

Страница 11: ...A This white circle is the light that is passing through the iris and it should be centered in the black field as in Figure 8b If it is not centered move it to the center using the centering screws iv...

Страница 12: ...Reduce the aperture iris diaphragm using the adjustment lever and rotate the lamp focusing control on the lamp housing until a sharp image of the lamp filament is attained 4 Refer to Figure 6 for thi...

Страница 13: ...ity filters as well as a clear opening 10 The Polarizing Analyzer is used in conjunction with the Polarized filter cube selected by pushing in the blue filter rod Figure 10 The Analyzer may also be pl...

Страница 14: ...from reducing the bulb s intensity and life 4 Reinsert the lamp socket into the lamp house and retighten the setscrew Make certain that the new lamp is placed squarely in the socket to insure that is...

Страница 15: ...on Be sure the instrument is dry before using 2 Cleaning optical parts Microscope eyepieces and objectives are coated They should not be wiped while dry as dirt or dust may scratch the coating It is b...

Страница 16: ...15 VWR TECHNICAL HOT LINE 888 VWR LINE 888 897 5463...

Отзывы: