background image

18 – VWR STandaRd Line PiPeTTORS 

instruction Manual – 19

 10.  Pipettor  Maintenance 

depending on the applications and intensity of use, 

the pipettor requires periodic maintenance. The com-

ponents exposed to corrosive vapors, such as shaft 

elements, should be regularly checked and cleaned.

do not use sharp tools for pipettor maintenance. 

it may cause damage to the device and affect 

the user’s safety.

Cleaning

external surfaces of the pipettor such as the push-

button, ejector pushbutton, handgrip, shaft nut, and 

adjustment knob may be cleaned using a cloth damp-

ened in isopropyl alcohol. The remaining parts removed 

from the pipettor during pipettor disassembly may be 

washed with distilled water or isopropyl alcohol.

NOTE

: Before using cleaning agents other than those 

recommended by the manufacturer, check the compat-

ibility charts and consider chemical resistance of the 

following plastics which form components of the pipet-

tor: PP, PC, POM, Pa, PPS, PVdF.

Sterilization 
Sterilization using an autoclave

The pipettor can be sterilized in an autoclave at 121°C 

for 20 minutes. Sterilization under other conditions may 

cause damage to the pipettor. it is recommended to:

 

Unscrew the shaft nut slightly in the 2-1,000 µL pipet-

tors, and unscrew the shaft slightly in the 5,000 µL and 

10,000 µL. after autoclaving these parts should be 

screwed tight again.

 

Sterilize the pipettors using an autoclave with an  

initial vacuum and drying cycle.

 

after sterilization, the pipettor should be dried and 

cooled to room temperature.

 

The shaft of the 5,000 µL and 10,000 µL models 

should be autoclaved without the filter.

Precision and accuracy should not alter if the pipet-

ting processes including autoclaving are carried out as 

described in this manual. if a change in accuracy occurs, 

it is recommended to:

 

Check the calibration of the pipettor after the first, 

third, and fifth autoclaving cycles and then after every 

10 autoclaving cycles.

Ultra Violet (UV) Sterilization

The pipettors are UV resistant. The distance from the 

radiation source to the exposed element of the pipet-

tor should be at least 50 cm. Prolonged or intense UV 

exposure can cause discoloration of pipettor parts but 

does not affect its performance.

 11. Troubleshooting 

if a problem is encountered during pipettor operation, 

use the following table to identify and eliminate the 

fault following the instructions provided. Replacement of 

parts should be required only occasionally and should 

not be required with normal pipettor use.

Problem

Cause

Solution

droplets 

of liquid 

remain in 

the pipet 

tip.

The pipet tip is emptied 

too fast.

decrease the speed 

of pressing 

the pipettor 

pushbutton.

The pipet tip wettability 

has increased due to 

extensive use.

Replace the tip with 

a new one.

droplets 

of air 

appear in 

the liquid 

aspirated 

into the 

pipet tip.

The pipet tip immersion 

depth is too shallow.

immerse the tip 

to the recommend-

ed depth according 

to the instructions.

The pipet tip is incor-

rectly pressed onto the 

pipettor shaft.

Press the pipettor 

tip firmly onto the 

pipettor.

The tip is damaged or 

worn out due to exten-

sive use.

Replace the tip with 

a new one.

The pipet-

tor 

incorrectly 

aspirates 

the liquid 

or the liquid 

drops out 

from the tip.

The pipettor tip is incor-

rectly pressed onto the 

pipettor shaft.

Press the pipet 

tip firmly onto the 

pipettor.

The shaft surface is 

damaged or contami-

nated at the sealing 

site.

Clean the shaft 

or replace it with 

a new one.

Содержание avantor

Страница 1: ...tandard Line Instruction Manual Pipettors VWR Standard Line VWR 8 channel Standard Line VWR 12 channel Standard Line This manual is available in additional languages at www vwr com PMS 647C CMYK 96 54...

Страница 2: ...s are designed for microplate filling The pipet tors enable precise and simultaneous delivery of 8 or 12 preset volumes of liquid These pipettors are available in four volume ranges 0 5 10 L 5 50 L 20...

Страница 3: ...ti channel manifold Contains a set of plungers and a set of flexible suspension shafts which func tion to reduce the force required to attach the tip Model Identification The volume range of the pipet...

Страница 4: ...Accuracy Precision Non filter Tips L Single channel Pipettors 2 Min 0 1 40 0 12 0 10 0 2 12 0 6 0 1 2 7 1 3 Max 2 1 5 0 7 10 Min 0 5 4 0 2 8 10 1 0 2 5 1 8 5 0 1 0 0 6 Max 10 0 0 5 0 4 20 Min 2 3 0 1...

Страница 5: ...o not attach the tips with a rocking movement as this may damage the shaft or plunger Observe this rule particularly with single channel pipettors of low volume range Never draw liquids directly into...

Страница 6: ...In 2 and 10 L pipettors place the cap that can be purchased separately on the bottom of the pipettor shaft and slide the cap upwards to the bottom of the tip ejector M L 10000 L L 10mm 5000 L L 6mm M...

Страница 7: ...ded to pre rinse the tip several times prior to use Do not pipette liquids with temperatures above 70 C NOTE When pipetting acids or corrosive solutions which emit vapors it is recommended to disassem...

Страница 8: ...ISO 8655 guidelines for the maximum nominal liquid volume drawn by the pipettor and for 10 of the maximum or minimum liquid volume according to the values given Section 5 The pipettors are designed t...

Страница 9: ...rm 10 aspirations and calculate the average value in mg Calculate the volume in L by multiplying the value in mg by the conversion factor Z L mg If the average aspirated volume exceeds the permissible...

Страница 10: ...accuracy should not alter if the pipet ting processes including autoclaving are carried out as described in this manual If a change in accuracy occurs it is recommended to Check the calibration of the...

Страница 11: ...ling elements are not sufficiently lubricated e g after repeated autoclaving procedures Incorrect aspiration Liquid with properties other than water den sity viscosity Calibrate the pipettor using the...

Страница 12: ...tional product information and special offers Contact us For information or technical assistance contact your local VWR representative or visit ww vwr com 14 Warranty VWR warrants that this product wi...

Страница 13: ...s Campus Northwest Business Park Ballycoolin Dublin 15 Tel 353 01 88 22 222 Email sales ie vwr com Italy VWR International S r l Via San Giusto 85 20153 Milano MI Tel 39 02 3320311 02 487791 Email inf...

Отзывы: