Gel Imagers 

 

12 

There is a 10-minute timer that turns off the transilluminator automatically after 10-minutes of use. The timer 
switch located at the front of the darkroom can be reset (which restarts the 10-minute count).  

 

Error Messages Pop Up on the Screen 
 

Messages appear on the screen that might be related to the Gel Imager software interface, or Microsoft 
Windows. If the message is related to Microsoft Windows, such as a reminder to activate or update your 
copy of Windows, please contact your system administrator for assistance. 

If an error message appears repetitively, and your system administrator does not recognize it as a Microsoft 
Windows error, contact UVP Technical Support for further assistance. 

 
Technical Support 

 

 . 

 

Warranty 

 

UVP's quality instruments are guaranteed to be free of defects in materials, workmanship, and manufacture 
for one (1) year from date of purchase. Transilluminators are similarly warranted for two (2) years from the 
date of purchase.  Consumable and disposable products, including, but not limited to lamps or light tubes, 
filters, or rechargeable batteries are guaranteed to be free from defects in manufacture and materials for 
ninety (90) days from date of purchase. If equipment failure or malfunction occurs during the warranty 
period, UVP shall examine the inoperative equipment and have the option of repairing or replacing any 
part(s) which, in the sole and absolute discretion of UVP, were originally defective or became so under 
conditions of normal usage and service. No warranty shall apply to any instrument, or part thereof, that has 
been subject to accident, negligence, alteration, abuse or misuse by the end user. Moreover, UVP makes no 
warranties whatsoever with respect to parts not supplied by UVP or that have been installed, used and/or 
serviced other than in strict compliance with the instructions appearing in this installation and instruction 
manual. In no event shall UVP be responsible to the end-user for any incidental or consequential damages, 
whether foreseeable or not, including but not limited to property damage, inability to use equipment, loss of 
business, loss of profits, or inconvenience arising out of or connected with the use of instruments produced 
by UVP. Nor is UVP liable or responsible for any personal injuries occurring as a result of the use, misuse, 
installation and/or servicing of equipment. Representations and warranties made by any person, including 
dealers and representatives of UVP which are inconsistent or in conflict with the terms of this warranty 
(including but not limited to the limitations of the liability of UVP as set forth above), shall not be binding 
upon UVP unless reduced to writing and approved by an expressly authorized representative of UVP.  THIS 
WARRANTY IS EXPRESSED IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, 
INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR 
PURPOSE AND OF ALL OTHER OBLIGATIONS OR LIABILITIES ON UVP’S PART AND UVP NEITHER 
ASSUMES NOR AUTHORIZES ANY PERSON TO ASSUME FOR UVP ANY OTHER LIABILITIES IN 
CONNECTION WITH THE SALE OF THE SAID PRODUCTS.   

 

Содержание 89131-380

Страница 1: ...Gel Imager and Gel Imager Plus Installation and User Instructions 81 0317 01 Rev A Gel Imager ...

Страница 2: ...chiving quality images in a compact standalone package No computer is required The models comes in two configurations Gel Imager Plus a high resolution 1 3 megapixel camera 8 48mm zoom lens or Gel Imager a 0 3 megapixel camera and a 5 7 34 2mm zoom lens System Components Components Refer to the packing slip and pictured components for parts included with the system Gel Imager Plus 8 LCD touch scre...

Страница 3: ... 89131 404 Gel Imager Plus 1 3 Camera 20 x 20cm Transilluminator 302 365nm UV 115V 89131 406 Gel Imager Plus 1 3 Camera 20 x 20cm Transilluminator 302 365nm UV 230V 89131 408 Gel Imager Plus 1 3 Camera 21 x 26cm Transilluminator 302 365nm UV 115V 89131 410 Gel Imager Plus 1 3 Camera 21 x 26cm Transilluminator 302 365nm UV 230V Camera System will be configured with one of the two camera options 1 3...

Страница 4: ...tors from models with 8 watt single wavelength and single intensity to models with multiple wavelengths and high low intensities Note For UV protection there is a 10 minute timer installed in the Gel Imager The timer shuts off the transilluminator after 10 minutes of operation A switch on the front of the darkroom allows the user to reset the timer Opening the darkroom door automatically resets th...

Страница 5: ...ite light box The Visi Blue Converter Plate converts UV to 460 470nm designed for use with blue excitation samples and SYBR green SYPRO orange and GFP stains Set up Instructions Components Camera with zoom lens bracket filter cables Gel Imager darkroom USB storage device 1 GB minimum Transilluminator WARNING DO NOT ATTEMPT TO CONNECT ANY WIRING WHILE THE EQUIPMENT IS CONNECTED TO ANY POWER SUPPLY ...

Страница 6: ...up when facing you Additional filters are available Refer to Accessories for ordering information Camera Set up The camera and zoom lens are assembled at the factory Note The zoon lens shipped may appear different than the one pictured at right Remove the cap from the lens Attach the step up ring and diopter to the lens The step up ring and diopter will only fit one way Using the four thumb screws...

Страница 7: ...to a surge protector power outlet Printer Set up The thermal printer is an optional accessory Proceed to the next section if a printer was not included with the shipment Stylus and Holder The screen is touch sensitive Use the stylus included with the system to operate the software controls For easy access to the stylus when operating the system remove the tape from the back of the holder and adher...

Страница 8: ...have power Turn on the transilluminator and darkroom power switches To power down the system push the main power button Adjusting the Screen Angle To adjust the angle of the screen loosen the knobs on each side of the monitor Twist each knob counter clockwise Note that the knobs will be turning opposite directions Tilt the screen to the desired angle Turn each knob clockwise to tighten ...

Страница 9: ... the Gel Imager darkroom power is turned on power is supplied to all components including the jumper power cable that supplies power to the transilluminator To use the transilluminator turn the transilluminator power on For UV protection there is a 10 minute timer installed in the system The timer shuts off the transilluminator after 10 minutes of operation A switch on the front of the darkroom al...

Страница 10: ...at turns the transilluminator off when the door is open After closing the door be sure the overhead white light switch is turned off While watching the LCD screen rotate the lens f stop adjustment so that the image is bright enough to see on the screen Rotate the lens focus adjustment on the lens Adjust this so that the image appears in clear focus on the LCD monitor Note Once the proper zoom rang...

Страница 11: ...the cap and the fuse holder will pop out Inspect the thin wire within the glass fuse to see if there is a break in the wire If so replace fuse s If fuses are blown repeatedly contact UVP Technical Support Department for additional troubleshooting No Image on the LCD Screen To power up the system push once on the button located on the left side of the unit Touch Screen not Accurate If touching or c...

Страница 12: ...acing any part s which in the sole and absolute discretion of UVP were originally defective or became so under conditions of normal usage and service No warranty shall apply to any instrument or part thereof that has been subject to accident negligence alteration abuse or misuse by the end user Moreover UVP makes no warranties whatsoever with respect to parts not supplied by UVP or that have been ...

Отзывы: