| Instruction manual
20
10.5 Adjusting interpupillary distance
Observing with both eyes, hold the two eyepiece prism
assemblies. Rotate them around their common axis until the
fields of view match.
The graduation on the interpupillary distance indicator (1),
pointed by the spot “.” on the eyepiece holder, shows the
distance between the operator’s eyes. (Fig. 24)
The range of the interpupillary distance is 48-75mm.
10.6 Use of eyeshields
Use with eyeglasses
Fold rubber eyeshields with both hands. Folded eyeshields avoid
scratching the lenses of eyeglasses. (Fig.
25)
Use without eyeglasses
Raise eye shields and observe at the microscope placing
eyes to the shields, avoiding external light to disturb the
observation. (Fig. 26)
10.7 Selecting the light path
Pull the light path selector lever (2) sideways using your thumb,
selecting the light path you need. (Fig. 27)
1
2
FIGURE 24
FIGURE 25
FIGURE 26
FIGURE 27
Light path lever
Brightness
Use
In
0% binocular observation /
100% video or photography
Microphotography or video
acquisition
Out
100% binocular observation /
0% video or photography
Binocular observation