Instruction Manual INCU-Line Cooled
Page
27
/
32
4.
On a regular basis clean with a vacuum cleaner of a duster the cooling unit and condenser
(exchanger) which is placed in the rear (or upper) part of the device.
Disobeying this pre-
caution may harm the compressor. In such case the warranty repair will not be ac-
cepted!
14.2 Interior cleaning
1.
The chamber should be emptied of any samples before cleaning.
2.
Open the door of the device and wait for the frost to melt, take out the shelves and start
cleaning the device,
3.
Only water or water with mild detergent should be used.
4.
Having finished cleaning, you should allow the device to dry completely.
5.
During cleaning pay attention not to damage the temperature sensor in the chamber.
15 RESUMING OPERATION AFTER A LONGER PERIOD OF TIME
1.
Remove all objects from the chamber.
2.
Disconnect the device from the mains.
3.
Clean and dry the chamber.
4.
Leave the door open to avoid bad smell.
5.
Store in temperature between 0
°C and 50°C and relative humidity maximum 70%.