10 STATISTICS
While a program is operating it is possible to check the average, minimum and maximum temperature val-
ue for all time-temperature segments. The values starts to register when unit reach the set temperature (or
to be more precise when the temperature inside the chamber is
0,2°C different from the programmed tem-
perature) at a ten-second interval.
Once a segment is finished, the recording stops and the device moves to the next temperature
– defined
for the next segment. When next set temperature is reached
– the recording starts again. For multi-
segment programs there are available statistics for each segment. If the program has not reached a seg-
ment yet, the display shows 'none data ' (there is no data available using which the average, minimum and
maximum values are calculated).
10.1 Preview of statistics.
In the main window press ESC
.
Enter STATISTICS using ENTER
.
Using
and
choose the segment which you want to see the
statistics of, then press ENTER
.
Using
and
you can see the average, max and min temper-
ature recordings.
You can also see how many recordings are available.
If the simple mode is active the statistics of chosen segment is not possible to preview.