• Samples should be aqueous. Measurements may be made in partially aqueous or some water-

miscible solvents. The results must be interpreted with caution as the full pH scale is shifted when
the solvent system changes.

• pH probes with an Ag/AgCl reference system and a single junction are not compatible with

solutions that contain silver complexing or binding agents such as TRIS, proteins and sulfides. To
measure in these solutions, use a double-junction pH probe, probe with an Ag

+

 (silver ion) barrier

or a Red Rod pH probe. Red Rod pH probes are compatible with these types of samples as they
have an encapsulated reference system.

• Proteins can collect on the sensing bulb. Make sure the probe stays clean when these types of

samples are measured.

• Do not use probes in solutions that are outside the temperature range of the probe.

Sample measurement

W A R N I N G

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current material safety data
sheets (MSDS) for safety protocols.

Before measurement:

Prepare the probe for use. Refer to 

Preparation for use

 on page 5.

Calibrate the probe. Refer to 

Calibration

 on page 6. The manufacturer recommends that the probe is calibrated

at least once a day for the best measurement accuracy.

Measurement notes

• The calibration and sample measurement conditions must be as similar as possible (e.g., the

temperature of the solution, stir procedure, stir rate and position of the probe).

• Calibration of conductivity cells is important as the conductivity cell constant can change with time

and calibration identifies the actual cell constant versus the nominal value.

• Do not dilute conductivity standards or samples.
• When the probe is submerged, make sure that there are no air bubbles under the probe sensor

tip(s). Gently shake the probe from side to side to remove any air bubbles.

• Make sure that the reference junction is fully in the solution.
• If necessary, turn the protective shroud and remove it from the probe. The protective shroud

prevents damage to the sensors.

• Do not put the probe on the bottom or sides of the container.
• If stabilization is slow, shake the probe from side to side in the solution.
• Do not use probes in areas where EMF is present (i.e., voltaic cells, thermoelectric devices,

electrical generators, resistors and transformers). For use in process units (i.e., spot checking),
make sure that the meter is grounded.

• If a measurement error occurs, refer to 

Troubleshooting

 on page 12.

English

   

9

Содержание 89231-656

Страница 1: ...DOC022 53 80319 sympHony Multisensor Probes 07 2012 Edition 1 User Manual...

Страница 2: ......

Страница 3: ...te For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items f...

Страница 4: ...or use with sympHony handheld meters only Product overview The multisensor probes are combination gel filled electrodes for pH conductivity and ORP or dissolved oxygen measurements in general aqueous...

Страница 5: ...e equipment to remove broken glass To prepare the probe for calibration or sample measurement 1 Turn the probe storage tube s and remove from the probe Keep the probe storage tube s 2 Rinse the probe...

Страница 6: ...pare fresh standard solutions for the calibration The calibration and sample measurement conditions must be as similar as possible e g the temperature of the solution stir procedure stir rate and posi...

Страница 7: ...onstant rate to prevent the formation of a vortex 6 Push Read Stir gently The display shows Stable when the reading is stable 7 When the reading is stable the meter prompts for the next calibration po...

Страница 8: ...dissolved oxygen calibration polarize the probe Refer to Polarize the dissolved oxygen probe 89231 660 multisensor on page 5 2 Push Calibrate and select the probe if applicable 3 Prepare the standard...

Страница 9: ...page 5 Calibrate the probe Refer to Calibration on page 6 The manufacturer recommends that the probe is calibrated at least once a day for the best measurement accuracy Measurement notes The calibrati...

Страница 10: ...surements are done prepare the probe for storage Refer to Storage on page 12 Maintenance Clean the probe N O T I C E Do not rub or touch the tip of the pH or dissolved oxygen sensors Clean the probe w...

Страница 11: ...fer to Fill the membrane module on page 11 Replace the membrane module N O T I C E Carefully handle the membrane modules to prevent damage It is necessary to replace the membrane module periodically R...

Страница 12: ...ge solution to the mark Refer to Specifications on page 3 for the applicable storage solution 2 Put the probe s in the storage tube s and turn to tighten Rehydrate the probe conductivity and pH sensor...

Страница 13: ...ectrode SAME BUFFERS Calibrate the probe Refer to Calibration on page 6 Examine the probe Refer to Examine the probe on page 13 Examine the standard solution Use a new standard solution Table 4 Measur...

Страница 14: ...14 English...

Страница 15: ......

Страница 16: ...alid in USA void where prohibited by law or company policy while supplies last Visit vwr com to view our policy and additional disclaimers VWR forms of VWR and the VWR logo and or design are either re...

Отзывы: