Refer to 

Table 3

 for calibration warnings and errors. Refer to 

Table 4

 for measurement warnings and

errors.

Table 3  Calibration warnings and errors

Error/Warning

Solution

Calibration out of range

The slope is out of range.
Calibrate the probe. Refer to 

Calibration

 on page 6.

Connect a new probe.

Standard or electrode in
poor condition

Calibrate the probe. Refer to 

Calibration

 on page 6.

Examine the probe. Refer to 

Examine the probe

 on page 13.

Examine the standard solution solution. Make sure that the standard solution used
is the standard solution selected in the calibration setup. Identify the temperature
specification in the calibration setup. Use a new standard solution.

Outside slope limits

Standard not recognized

Unstable reading Time t >
120 s

Calibrate the probe. Refer to 

Calibration

 on page 6.

Examine the probe. Refer to 

Examine the probe

 on page 13.

Make sure that the probe is correctly immersed in the sample.

Electrode in poor condition

Examine the probe. Refer to 

Examine the probe

 on page 13.

Check the electrode

SAME BUFFERS

Calibrate the probe. Refer to 

Calibration

 on page 6.

Examine the probe. Refer to 

Examine the probe

 on page 13.

Examine the standard solution. Use a new standard solution.

Table 4  Measurement warnings and errors

Error/Warning

Solution

pH 12.78 19°C

Calibrate the probe. Refer to 

Calibration

 on page 6.

Measurement out of range
("_ _ _ _ _ _")

Examine the probe. Refer to 

Examine the probe

 on page 13.

Unstable reading

Examine the probe. Refer to 

Examine the probe

 on page 13.

Make sure that the probe is correctly immersed in the sample.

The temperature is out of °C
range.

Examine the temperature sensor.
Connect a different probe to identify if the problem is with the probe or the
meter.

Time > 120 s

Make sure that the probe is correctly immersed in the sample.
Measure the sample temperature.
Examine the probe. Refer to 

Examine the probe

 on page 13.

Time > 240 s

Time > 300 s

Examine the probe

Note: The lower the sample temperature or the larger the temperature difference between the samples, the longer
the stabilization time will be.

1.

Clean the probe. .

2.

Turn off the meter. Disconnect and then connect the probe again.

3.

Connect a different probe to identify if the problem is with the probe or the meter.

English

   

13

Содержание 89231-656

Страница 1: ...DOC022 53 80319 sympHony Multisensor Probes 07 2012 Edition 1 User Manual...

Страница 2: ......

Страница 3: ...te For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items f...

Страница 4: ...or use with sympHony handheld meters only Product overview The multisensor probes are combination gel filled electrodes for pH conductivity and ORP or dissolved oxygen measurements in general aqueous...

Страница 5: ...e equipment to remove broken glass To prepare the probe for calibration or sample measurement 1 Turn the probe storage tube s and remove from the probe Keep the probe storage tube s 2 Rinse the probe...

Страница 6: ...pare fresh standard solutions for the calibration The calibration and sample measurement conditions must be as similar as possible e g the temperature of the solution stir procedure stir rate and posi...

Страница 7: ...onstant rate to prevent the formation of a vortex 6 Push Read Stir gently The display shows Stable when the reading is stable 7 When the reading is stable the meter prompts for the next calibration po...

Страница 8: ...dissolved oxygen calibration polarize the probe Refer to Polarize the dissolved oxygen probe 89231 660 multisensor on page 5 2 Push Calibrate and select the probe if applicable 3 Prepare the standard...

Страница 9: ...page 5 Calibrate the probe Refer to Calibration on page 6 The manufacturer recommends that the probe is calibrated at least once a day for the best measurement accuracy Measurement notes The calibrati...

Страница 10: ...surements are done prepare the probe for storage Refer to Storage on page 12 Maintenance Clean the probe N O T I C E Do not rub or touch the tip of the pH or dissolved oxygen sensors Clean the probe w...

Страница 11: ...fer to Fill the membrane module on page 11 Replace the membrane module N O T I C E Carefully handle the membrane modules to prevent damage It is necessary to replace the membrane module periodically R...

Страница 12: ...ge solution to the mark Refer to Specifications on page 3 for the applicable storage solution 2 Put the probe s in the storage tube s and turn to tighten Rehydrate the probe conductivity and pH sensor...

Страница 13: ...ectrode SAME BUFFERS Calibrate the probe Refer to Calibration on page 6 Examine the probe Refer to Examine the probe on page 13 Examine the standard solution Use a new standard solution Table 4 Measur...

Страница 14: ...14 English...

Страница 15: ......

Страница 16: ...alid in USA void where prohibited by law or company policy while supplies last Visit vwr com to view our policy and additional disclaimers VWR forms of VWR and the VWR logo and or design are either re...

Отзывы: