Measurement procedure

1. 

 Connect the probe

to the meter. Turn the
meter on.

Note: Before dissolved
oxygen measurement,
polarize the probe if the
probe is not polarized.

2. 

 Rinse the probe with

deionized water. Blot
dry with a lint-free cloth.

3. 

 Put the probe in the

sample and stir gently
at a constant rate to
prevent the formation of
a vortex.

4. 

 Push 

Read

. Select

the channel for the
parameter to be
measured if applicable.

5. 

 Stir gently. The

display shows "Stable"
when the reading is
stable.

6. 

 If the folder icon is

shown, push the folder
icon to save the data.

Note: Data is
automatically saved if
the folder icon is not
shown.

7. 

 Do steps 

2

6

 again

for additional
measurements.

8. 

 When

measurements are
done, prepare the
probe for storage. Refer
to 

Storage

on page 12.

Maintenance

Clean the probe

N O T I C E

 

Do not rub or touch the tip of the pH or dissolved oxygen sensors.

Clean the probe when there is contamination on the sensor(s). Symptoms of contamination are:

• Readings are not accurate or consistent.
• The stabilization time is slow.
• A calibration error occurs.
• Contamination is visible on the probe (i.e., dirt).

To clean the dissolved oxygen probe of the multisensor probe, refer to 

Dissolved oxygen probe

(89231-660 multisensor)

 on page 11.

1.

For general contaminants, rinse the probe with deionized water. Blot dry with a lint-free cloth.

2.

For other contaminants:

a.

Soak the pH and/or conductivity sensors of the probe in the applicable cleaning agent. Refer
to 

Table 2

.

b.

Rinse or soak the sensors for 1 minute in deionized water. Blot dry with a lint-free cloth.

10

   

English

Содержание 89231-656

Страница 1: ...DOC022 53 80319 sympHony Multisensor Probes 07 2012 Edition 1 User Manual...

Страница 2: ......

Страница 3: ...te For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items f...

Страница 4: ...or use with sympHony handheld meters only Product overview The multisensor probes are combination gel filled electrodes for pH conductivity and ORP or dissolved oxygen measurements in general aqueous...

Страница 5: ...e equipment to remove broken glass To prepare the probe for calibration or sample measurement 1 Turn the probe storage tube s and remove from the probe Keep the probe storage tube s 2 Rinse the probe...

Страница 6: ...pare fresh standard solutions for the calibration The calibration and sample measurement conditions must be as similar as possible e g the temperature of the solution stir procedure stir rate and posi...

Страница 7: ...onstant rate to prevent the formation of a vortex 6 Push Read Stir gently The display shows Stable when the reading is stable 7 When the reading is stable the meter prompts for the next calibration po...

Страница 8: ...dissolved oxygen calibration polarize the probe Refer to Polarize the dissolved oxygen probe 89231 660 multisensor on page 5 2 Push Calibrate and select the probe if applicable 3 Prepare the standard...

Страница 9: ...page 5 Calibrate the probe Refer to Calibration on page 6 The manufacturer recommends that the probe is calibrated at least once a day for the best measurement accuracy Measurement notes The calibrati...

Страница 10: ...surements are done prepare the probe for storage Refer to Storage on page 12 Maintenance Clean the probe N O T I C E Do not rub or touch the tip of the pH or dissolved oxygen sensors Clean the probe w...

Страница 11: ...fer to Fill the membrane module on page 11 Replace the membrane module N O T I C E Carefully handle the membrane modules to prevent damage It is necessary to replace the membrane module periodically R...

Страница 12: ...ge solution to the mark Refer to Specifications on page 3 for the applicable storage solution 2 Put the probe s in the storage tube s and turn to tighten Rehydrate the probe conductivity and pH sensor...

Страница 13: ...ectrode SAME BUFFERS Calibrate the probe Refer to Calibration on page 6 Examine the probe Refer to Examine the probe on page 13 Examine the standard solution Use a new standard solution Table 4 Measur...

Страница 14: ...14 English...

Страница 15: ......

Страница 16: ...alid in USA void where prohibited by law or company policy while supplies last Visit vwr com to view our policy and additional disclaimers VWR forms of VWR and the VWR logo and or design are either re...

Отзывы: