4/ condensation collecting on CO2 sensor
5/ CO2 sensor or interface failure
6/ unit incorrectly calibrated, see section 5.11
8/ taking Fyrite reading too soon after the door has been opened
9/ air leak around CO2 sensor mounting plate
MECHANICAL
Door not sealing
1/ check physical condition of gasket
2/ confirm that door latch pulls door in tightly
3/ assure that gasket is in original location
Motor doesn't move
1/ if shaft spins freely: check connections to motor and check
voltage to motor;
2/ if shaft rubs or is frozen, relieve binding and retest
Motor makes noise
1/ If noise is from the motor, tap the top of motor shaft with ball
peen hammer.
2/ If the sound gets worse, tap the other end of the shaft -
avoiding touching the fan blade.
3/ If there is no change, call Customer Service.
4/ If noise is from shaft or fan blade, realign shaft.
Solenoid valve buzzing
After removing solenoid clean with alcohol for carbon build-up
then blow out. Check valve seat or channels for contamination.
Check CO2 filter and/or grade of CO2 used. If seat is worn,
replace.
OTHER
Controller on at all times - "locked-up"
1/ Adjust set point to room temperature. If the light goes out but is
still heating, replace the solid state relay.
2/ turn unit off and on to reset
3/ if cannot change any condition on the front panel, call Customer
Service
Front panel displays are all off
1/ Check for wire damage.
unit or wall fuse/circuit breaker is
blown
1/ check wall power source
2/ compare current draw and compare to specs on data plate
3/ see what other loads are on the wall circuit
unit will not turn on
1/ check wall power source
2/ check fuse/circuit breaker on unit or in wall
3/ see if unit is on, e.g., fan or heater, and just controller is off
4/ check all wiring connections, esp. around the on/off switch
Unit is smoking – Out of box
Put unit under vent and run at full power for one hour.
Contamination in chamber
1/ see cleaning procedure in operator’s manual
2/ develop and follow Standard operating procedure for specific
application; include definition of cleaning technique and
maintenance schedule.