Display and Fyrite reading do not
match
1/ calibration error - clear chamber for 12 hrs and confirm at "0"
2/ turn set point up and down to see if solenoid valve works by
feeling and listening to valve
3/ assure that IR tube block is completely tight
4/ if display is varying when doing Fyrite test, have a leak on IR
base or mounting plate, or CO2 is injecting at time of test.
5/ take voltage reading across sensor leads 0=0V and 20=1V
6/ assure that sensor is properly mounted
7/ disp lower than fyrite: air leak around the CO2 sensor or fan
motor shaft
8/ display higher than fyrite: leak in tube from sample port to
chamber
9/ display higher than fyrite: Fyrite fluid worn out, replace fluid with
Fyrite refill.
10/ display higher than fyrite: poor Fyrite sampling technique, see
section 6.0
11/ display higher than fyrite: defective Fyrite pump or tubing
12/ CO2 sensor, interface or controller failure
Is unstable – display or actual reading
varies around set point
1/ confirm that fan is working
2/ check for air leak around CO2 sensor or fan motor shaft
3/ magnetic field interference
4/ poor door seal
5/ atmospheric pressure fluctuations
6/ defective Fyrite pump or tubing
7/ CO2 entering chamber too quickly
a- tank pressure too high
b- regulator set wrong
c- restrictor incorrect or missing
8/ electronic problem with CO2 sensor, interface or controller
9/ top of unit exposed to cold air drafts
10/ unit being operated without shadow box cover
11/ electrical interference – atmospheric or conducted
12/ incubator too heavily loaded
Can't adjust set points or calibration -
"locked up"
1/ turn unit off, then on to reset processor in controller
2/ if repeatedly happens, call Customer Service
Feeding continuously or abnormally
high CO2 usage
1/ do decay test: if more than 1% decay in an hour, check for leak
: door gasket tightness, motor inlet to chamber, sensor and
probe inlet to chamber
2/ check output signal from controller
3/ check solenoid valve for correct operation
4/ leak in plumbing including between regulator and CO2 tank
5/ door being opened too often
Won’t hold calibration – Fyrite reading
varies but display stable
1/ atmospheric pressure fluctuations
2/ top of unit exposed to cold air drafts
3/ unit being operated without shadow box cover in place