- 38 -
CAL:VOLT:DATA {P1|P2},{DATA}
This command revises voltage parameter from selected load with the data read from
voltmeter.
ex.)
CAL:VOLT:DATA
P1,”12.345”
[Warning]
You must enter minimum and maximum parameter.
Voltage Revising Procedure
(1) CAL:STAT ON,1234 Please unlock firewall protection.
(2) CAL:VOLT:LEV P1,MIN Check the lowest voltage parameter
(3) CAL:VOLT:DATA P1,0.004
Type the lowest voltage parameter from voltmeter
(4) CAL:VOLT:LEV P1,MAX
Check the highest voltage parameter
(5) CAL:VOLT:DATA P1,30.002 Type the highest voltage parameter from voltmeter
(6) CAL:STAT OFF,1234
Get out from setup screen
CAL:CURR:LEV {P1|P2},{MIN|MAX}
This command sets current parameter level.
ex.) if you type
“
CAL:CURR:LEV P1,MIN
” you will see the lowest current parameter from
load 1.
CAL:CURR:DATA {P1|P2},DATA
This command revises current parameter from selected load with the data read from
amperemeter
.
ex.)
CAL:CURR:DATA
P1,”0.545”
[Warning]
You must enter minimum and maximum parameter.
Current Revising Procedure
(1) CAL:STAT ON,1234 Please unlock firewall protection.
(2) CAL:CURR:LEV P1,MIN Check the lowest current parameter
(3) CAL:CURR:DATA P1,0.002
Type the lowest current parameter from
amperemeter.
(4) CAL:CURR:LEV P1,MAX
Check the highest current parameter
(5) CAL:CURR:DATA P1, 4.999 Type the highest current parameter from
amperemeter.
(6) CAL:STAT OFF,1234
Get out from setup screen