- 29 -
4-2. Current Calibration
Order
Selection
Screen Display
Description
1
P1 or P2
SET-Limit
SET-CAL_Current
Select desired output port.
Press
“SETUP” button.
Please check whether OVP or OCP
function is turned OFF.
Use Arrow sign or Jog Shuttle to choose
SET-Cal_Current.
2
SET-CAL_Current
Password **** 0
※
Factory default
- 1 2 3 4
Press
“SETUP” button.
Please enter 4 digit Password.
Use the arrows to move cursor.
Use the jog shuttle to select the number
for the password.
3
SET-CAL_Current
Alo
– 0.000A
By pressing
“STORE” button, you will
be able to restore lowest current level.
Enter 3 decimal point current value that
ampere meter shows.
Use the arrows to move cursor.
Use the jog shuttle to select the
number.
4
SET-CAL_Current
Wait For AD Cal
SET-CAL_Current
Ahi
– 5.000A
Press
“STORE” Button to save the
lowest current level.
After 2 seconds, the screen will
automatically lead you to the highest
current level selection.
Enter 3 decimal point voltage value that
ampere meter shows.
Use the arrows to move cursor.
Use the jog shuttle to select the
number.
5
Wait For AD Cal
Complete
P1 00.000V 0.000A
Press
“STORE” button to save the
restored value
The unit will be restored after 2
seconds.
4. Cailbration