- 18 -
3-7. Over Current Protection (OCP)
This feature turns off the output when the current exceeds the OCP (Over Current Protection) value set
by the user, and protects Load (DUT) connected to the output terminals. Over Current Protection
operates normally within the value range specified in the
“Specifications” section of this manual.
[ IMPORTANT ]
OCP should be set to
“OCP OFF" during unit calibration.
○
OCP Settings
No.
Key
Operation
Display Panel
Description
1
P1 or P2
** OUTPUT OFF **
SET-Limit
SET-OCP
Select the output port to use.
Turn off output by pressing
“OUTPUT
ON / OFF
” button (recommended).
Press
“SETUP” button.
Use the arrows or jog shuttle to select
“SET-OCP”.
2
SET-OCP
OCP ON / OFF
**OUTPUT OFF **
(If
“OCP OFF” is selected)
**OCP ON**
(If
“OCP ON” is selected)
SET-OCP
Level - 00.000A
Press
“SETUP” button again.
Use the jog shuttle to slect ON or OFF
Press
“STORE” button to save the set
value. If
“OCP OFF” is selected, it will
turn off OCP and exit Setup.
If
“OCP ON” is selected, user can set the
OVP Level.
3
SET-OCP
Level - 00.000V
** OUTPUT OFF **
OCP
P1 00.000V 0.000A
OCP
Use the arrows to change the location of
the cursor and use the jog shuttle to set
the value.
Press the
“STORE” button.
It will save the set value and exit Setup.
If OCP is on, it will indicate
“OCP” in the
display panel.
Press
“OUTPUT ON / OFF” button again
to continue use.
3. Features