EX2500A Web Page Operation
59
LXI
S
YNCHRONIZATION
The
LXI Synchronization
web page provides status on the EX2500A’s synchronization status.
F
IGURE
5-8:
LXI
S
YNCHRONIZATION
W
EB
P
AGE
•
IEEE 1588 Parameters
o
PTP Grandmaster Clock MAC
: indicates the MAC address of the device on the
network to which all other IEEE 1588 compliant devices are synchronized.
o
PTP Parent Clock MAC
: indicates the MAC address of the device to which the
EX2500A is synchronized. On a LAN that consists of multiple subnets, this indicates the
MAC address of the parent clock on that particular subnet.
o
PTP State
: indicates whether the EX2500A is a master, slave, faulty, disabled, passive,
or unlocked.
o
Current PTP Time
: indicates the number of seconds since midnight January 1, 1970
UTC (coordinated universal time).
o
Current Local Time
: indicates the date and time (in 24-hour format).
o
Grandmaster Traceability to UTC
: indicates how the grandmaster is synchronized to
the UTC. The values provide an indicator of how closely synchronized the grandmaster is
to UTC time.
Value
Clock Source
UTC Synchronization
ATOM
UTC Atomic Clock
Less than 100 ns
GPS
GPS Receiver
Less than 100 ns
NTP
Network Time Protocol
Less than 50 ms
HAND
Manually/Automatically Set
Less than 10 s
INIT
User Defined
Not specified
DFLT
None
None
T
ABLE
5-1:
UTC
G
RANDMASTER
C
LOCK
I
DENTIFIERS
o
Current Variance of Parent
: this function not currently supported by the EX2500A.
o
IEEE 1588 Domain
: indicates the instrument’s PTP subdomain. Typically, this indicator
is _DFLT.